Performance Veri
fi
cation
14.
If necessary, turn the Horizontal
Position
knob to set the delay to exactly
1.0000 ms
.
15.
Compare the rising edge of the marker with the center horizontal graticule
line. The rising edge should cross the 0 V center within ±2.5 divisions
(±25 ns) of the center graticule line. Enter the deviation in the test record.
NOTE.
One division of displacement from graticule center corresponds to a
10 ppm time base error.
Check Digital Threshold
Accuracy (MSO2000 Series
only)
For the MSO2000 series only, this test checks the threshold accuracy of the digital
channels. This procedure applies to digital channels D0 through D15, and to
channel threshold values of 0 V and +4 V.
1.
Connect the P6316 digital probe to the MSO2000 series instrument.
2.
Connect one of the digital channels, such as D0, to the DC voltage source to
run this test.
If using the Fluke calibrator as the DC voltage source, connect the calibrator
head to the digital channel to test. You will need a BNC-to-0.1 inch pin
adapter to complete the connection. Be sure to connect the digital channel to
the corresponding signal pin and to a ground pin on the adapter.
3.
Push the front-panel
Default Setup
button to set the instrument to the factory
default settings.
4.
Push the front-panel
D15-D0
button.
5.
Push the
D15-D0 On/Off
lower-bezel button.
6.
Push the
Turn On D7 - D0
and the
Turn On D15 - D8
side-bezel buttons.
The instrument will display the 16 digital channels.
7.
Push the
Thresholds
lower-bezel button.
26
DPO2000 and MSO2000 Series Speci
fi
cations and Performance Veri
fi
cation