Performance Veri
fi
cation
4.
Push the
Source
lower-bezel button.
5.
Select the
AC Line
trigger source with Multipurpose knob
a
. You do not need
to connect an external signal to the oscilloscope for this DC Balance test.
6.
Push the front-panel
Acquire
button.
7.
Push the
Average
lower-bezel button, and then push the
Average
side bezel
button to turn averaging ON.
NOTE.
When using averaging, allow the oscilloscope to acquire all the samples
before taking the measurement.
8.
If needed, adjust the number of averages to
16
with Multipurpose knob
a
.
9.
Push the front-panel channel button for the oscilloscope channel to test, as
shown in the test record (for example, 1, 2, 3, or 4).
10.
Set the channel being tested to 200 mV/div using the Vertical
Scale
knob.
11.
Attach a 50
Ω
terminator to the oscilloscope input channel being tested.
12.
Push the lower-bezel
Coupling
button to select
DC
or
GND
coupling, as
given in the test record.
13.
Push the front-panel Wave Inspector
Measure
button.
14.
Push the
Add Measurement
lower bezel button.
15.
Use Multipurpose knob
a
to select the
Mean
measurement.
16.
Push the
OK Add Measurement
side-bezel button, and then push the
Menu
Off
button.
17.
View the mean measurement value in the display and enter that mean value
as the test result in the test record.
18.
Repeat step 9, then steps 12 through 17.
NOTE.
When using averaging, make sure to allow the oscilloscope to acquire all
the samples before taking the measurement.
19.
Repeat steps 9 through 18 for each remaining channel.
DPO2000 and MSO2000 Series Speci
fi
cations and Performance Veri
fi
cation
19