Operating Requirements
TPP0100 or a TPP0200
Passive Probe
Single-ended voltage probe (ground referenced): 300 V
RMS
CAT II safety requirements
Installation Category II - for measurements performed on circuits directly connected to the
low-voltage installation
Temperature:
Operating: -10 °C to +55 °C (+14 °F to +131 °F)
Nonoperating: -51 °C to +71 °C ( -60 °F to +160 °F)
Pollution Degree: 2, Indoor use only
Humidity: 5% to 95% RH
MSO2000B Series
Oscilloscope with a P6316
Digital Probe
Threshold Accuracy: ±(100 mV + 3% of threshold)
Threshold Range: ±20 V
Maximum nondestructive input signal to probe: ±40 V peak
Minimum signal swing: 500 mV
peak-to-peak
Input resistance: 101 k
Ω
Input capacitance: 8.0 pF
Temperature:
Operating: 0 °C to +50 °C (+32 °F to +122 °F)
Nonoperating: -20 °C to +60 °C (-4 °F to +140 °F)
Altitude:
Operating: 3,000 m (9,843 ft) maximum
Nonoperating: 12,000 m (39,370 ft) maximum
Pollution Degree: 2, Indoor use only
Humidity:
5% to 95% relative humidity
MSO2000B and DPO2000B Installation and Safety Instructions