DMM800 Series Performance Verification
18
Handheld Instruments Basic Service
4.
Turn the calibrator output off.
5.
Disconnect the calibrator from the multimeter.
Perform the following steps to verify the AC ampere measurement accuracy.
1.
Set the multimeter dial to A .
2.
Push the BLUE button to select AC mode.
3.
Connect the calibrator outputs to the multimeter A and COM input
connectors.
4.
Set the calibrator to each of the values in the AC ampere test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
5.
Turn the calibrator output off.
6.
Disconnect the calibrator from the multimeter.
AC Ampere Check