DMM800 Series Performance Verification
Handheld Instruments Basic Service
17
4.
Set the calibrator to each of the values in the Volts peak hold test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
5.
Turn the calibrator output off.
6.
Disconnect the calibrator from the multimeter.
Perform the following steps to verify the DC milliampere measurement accuracy.
1.
Set the multimeter dial to
�
A or mA as needed.
2.
Connect the calibrator outputs to the multimeter
�
A mA and COM input
connectors.
3.
Set the calibrator to each of the values in the DC milliampere test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
4.
Turn the calibrator output off.
5.
Disconnect the calibrator from the multimeter.
Perform the following steps to verify the AC milliampere measurement accuracy.
1.
Set the multimeter dial to
�
A or mA as needed.
2.
Push the BLUE button to select AC mode.
3.
Connect the calibrator outputs to the multimeter
�
A mA and COM input
connectors.
4.
Set the calibrator to each of the values in the AC milliampere test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
5.
Turn the calibrator output off.
6.
Disconnect the calibrator from the multimeter.
Perform the following steps to verify the DC ampere measurement accuracy.
1.
Set the multimeter dial to A .
2.
Connect the calibrator outputs to the multimeter A and COM input
connectors.
3.
Set the calibrator to each of the values in the DC ampere test record and
verify that the multimeter reads within the specified Display minimum and
maximum limits.
DC Milliampere Check
AC Milliampere Check
DC Ampere Check