Performance veri
fi
cation
Basic mode: Spurious test
This test veri
fi
es the spurious using a spectrum analyzer. The generator should be
set to Basic mode operation.
1.
Connect the generator to the spectrum analyzer as shown in the following
fi
gure.
Figure 12: Spurious tests
2.
Set up the arbitrary function generator using the following steps:
Select menu
Setting
Operation
Function
Sine
Home > Sine
Frequency
5 MHz
Home > Freq
Amplitude Units
Vpp
Home > Units> Vpp
Amplitude
1.00 Vpp
Home > Ampl
Channel 1 Output
On
On (front)
3.
Set the center frequency of the spectrum analyzer to 10 MHz. Other settings
are shown in the following tables.
4.
Set the Ref Level of the spectrum analyzer to 10 dBm.
5.
Measure the maximum spurious level other than harmonic distortion of 1 Vpp
sine wave in each frequency.
6.
Set the center frequency of the spectrum analyzer to 625 MHz. Other settings
are shown in the following tables.
7.
Measure the maximum spurious level other than harmonic distortion of 1 Vpp
sine wave in each frequency.
8.
Verify that the spurious signal at each frequency is equal to or less than the
limit speci
fi
ed in the following tables.
9.
Record the results in the test record.
10.
Repeat steps 2 through 9 for the CH1–, CH2+, and CH2– outputs.
42