Speci
fi
cations
All speci
fi
cations are guaranteed unless labeled “typical”. Typical speci
fi
cations
are provided for your convenience, but are not guaranteed.
Speci
fi
cations that are check marked with the
symbol are checked directly in
the Performance Veri
fi
cation section.
All speci
fi
cations apply to the arbitrary function generator unless noted otherwise.
These speci
fi
cations are valid under the following conditions:
The instrument must have been calibrated/adjusted at an ambient temperature
b20 °C and +30 °C.
The instrument must be operating at an ambient temperature between 5 °C
and +50 °C.
The instrument must have had a warm-up period of at least 30 minutes.
The instrument must be in an environment with temperature, altitude, and
humidity within the operating limits described in these speci
fi
cations.
Electrical speci
fi
cations
Table 1: Clock generator
Characteristic
Description
Sampling rate control
100 S/s to 2.5 GS/s
Resolution: 8 digits; 100 S/s
2 channels can be set to different sampling rates. Each channel’s sampling clock is individual.
Internal clock stability
±0.28 x 10
-6
(Frequency stability vs. change in temperature.)
Holdover stability: ±0.32 x 10
-6
(Inclusive of frequency stability, supply voltage changes (±1%),
aging, for 24 hours.)
Constant temperature stability: ±40 x 10
-9
Internal clock aging, typical
Per life (20 years): within ±3 x 10
-6
Per day: within ±40 x 10
-9
Reference oscillator error
(cumulative)
Total reference frequency error, one year after factory adjustment, at any temperature within
operating limits.
±1.6 x 10
-6
/year
Includes allowances for Aging per Year, Reference Frequency Calibration Accuracy, and
Temperature Stability.
Aging Per Year: ±1.0 x 10
-6
Reference oscillator calibration
accuracy
±1.0 × 10
-6
when operated within 25 °C ±5 °C, after 30 minute warm-up
Reference oscillator aging
Within ±1.0 x 10
-6
/year
AWG4162 Speci