7.
METER
a.
Meter zero
CALIBRATION
Set TEST power to off and INDICATION to HTR. Ad
just meter zero adjustment for an indicated zero.
Reset INDICATION to .+DC.
b.
Meter accuracy
error: ±2%. max
Connect voltmeter between test panel DC+ jack and
gnd. Check 570 meter for proper dc voltage reading
( ±2%) by comparing against test meter as follows:
(1)
Set VADC fully ccw.
(2)
Set +DC to 200
(3)
Set TEST power to ON.
( 4)
Adjust VARIABLE +DC for +140:, as read on
test meter.
(5)
Read 570 meter: 14 0v, ±2%. max.
(6)
Set +DC to 100 and VARIABLE +DC to CALI-
BRATED.
,
(7)
Read. 570 meter: 100 v. ±2%. max .
12
570 CALIBRATION
NOTES
2-5 -64
TEK 570
FC'.D
Содержание 570
Страница 2: ...Verso Filler Page ...
Страница 6: ...Verso Filler Page ...
Страница 11: ......
Страница 12: ...Verso Filler Page ...
Страница 16: ......
Страница 42: ...NOTES ...
Страница 77: ......
Страница 78: ...Verso Filler Page ...
Страница 79: ......
Страница 80: ...Verso Filler Page ...
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......
Страница 90: ...Verso Filler Page ...
Страница 91: ......
Страница 92: ...Verso Filler Page ...
Страница 93: ......
Страница 94: ...Verso Filler Page ...
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ...Verso Filler Page ...
Страница 134: ...Verso Filler Page ...