![Tektronix 535A Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/tektronix/535a/535a_instruction-manual_1078931066.webp)
Calibration
Procedure
—
Type
535A
29.
Set
Lockout Level Adjust
Set
the HORIZONTAL DISPLAY switch at A'
DEL'D BY
B’
the B'
STABILITY control fully clockwise, and the A'
TIME/CM
switch at .1 MILLISEC.
Slowly
adjust
the
A'
STABILITY
control until the sweep first appears. Connect
the
test
oscilloscope through a 10X probe to pin 3 of V133
using
dc
coupling. Adjust the test oscilloscope so that the
displayed
waveform has a vertical amplitude
of four divi
sions.
Adjust
the
Lockout Level Adj. control (see top view,
Fig.
6-3) until the square-wave portion of the displayed
waveform
is 2.4 divisions in amplitude. During adjustments,
the A'
STABILITY control should be checked frequently to
be
sure that it is set where the sweep just runs. At the
completion
of
the adjustment, the
square wave portion
should
be
2.4
centimeters in amplitude, the
sawtooth portion
should
be
1.6 centimeters
in amplitude.
30.
Set
Vertical Gain
Connect
the EP53A Gain Set Adaptor between the oscillo
scope
and plug-in unit. Connect
a
jumper from the CAL.
OUT
connector to the connector
of the gain set adaptor.
Place
the AMPLITUDE CALIBRATOR switch in the .2 VOLTS
position and adjust the
oscilloscope
for a
stable display.
Set
the Gain Adj.
(see Figure 6-5) control for
exactly 2 centi
meters of
vertical
deflection.
31.
Adjust
Delay Line
and High Frequency
Compensations
Delay Line
adjustments are probably the most difficult
adjustments
to
make. This
is
due
to interaction
between the
large number
of
controls.
The
adjustments can
be made
properly
with
a minimum of difficulty if the proper pro
cedure is
followed
and
if care is used in making the adjust
ments.
In
general, it is not necessaary to completley re
adjust
the delay line. Rather it
is a case of occasionally
touching
up the line
for best results. Before attempting to
adjust
the
line, briefly run through this procedure to
estab
lish the
basic concepts in your
mind. This will make the
actual adjustments
considerably easier.
Be sure that the
line
requires adjustment
before attempting to adjust it.
In
general the procedure for adjusting the delay
line and
high
frequency controls involves application of a fast ris
ing pulse
or square wave to the vertical amplifier. The
delay
line is then adjusted for minimum distortion of the
pulse wave
displayed on the screen. In
particular, the
delay
line and amplifier
high frequency controls are adjusted
for
three characteristics of the displayed waveform; fastest
possible
risetime
without overshoot,
minimum irregularities
on
the top
of
the pulse, and minimum slope of the top of
the
waveform.
Displaying
the Test Signal
To
determine the extent of misadjustment of the delay line
in
your
instrument,
you will need
to
closely examine a dis
played
fast rising pulse or square wave.
The
waveform
used
to make this examination should have a risetime of no
more
than
7
millimicroseconds.
A
Tektronix
Type 107
Square-Wave
Generator or a Tektronix Type P Plug-In Unit
is
recommended.
If
a
Type
107 Square-Wave Generator is used, the fre
quency
control should be
set to obtain
approximately 400 kc
squarewaves.
The
output of the generator should be con
nected
through
a 52 ohm coaxial cable
and
termination
resistor to
the input of either a Type
K or a Type L Plug-In
Unit.
The
termination
resistor should be connected at the
input to the
plug-in unit.
When
a Type P Plug-In Unit is used, it is only necessary
to
insert the
unit into
the plug-in compartment of the oscil
loscope.
The pulse frequency of the P Unit is fixed.
If
it
is necessary
to use
a signal generator other than a
Type
107 or
Type P,
you must check the output waveform
from
the generator for suitability. Obviously a waveform
which
is
distorted in any manner is not
suitable for adjust
ing
the delay line of an oscilloscope. A good check on the
suitability
of your test equipment is to display the output
waveform
on another Tektronix
oscilloscope with a delay line
known
to be correctly adjusted.
When
the input
signal has been connected, adjust the
oscilloscope
triggering controls
for a stable
display
using
Time
Base
A. Rotate the TIME/CM control to .1
µ
SEC and
adjust
the oscilloscope and signal
source for
approximately
3 centimeters
of vertical
deflection.
With
the
oscilloscope display, it is possible to check the
delay
line
and amplifier adjustments for proper setting.
There are
three general characteristics
which you will have
to
appraise, and to do this, you will
need to use three dif
ferent
sweep
rates.
The
first characteristic to look for is
the
level
of the display; the second is the amount of bump
iness
contained in the flat top of the displayed waveform;
and
the third is the risetime of the displayed waveform.
Determining
the
Level of the
Display
The
level of the display refers to the top of the displayed
waveform.
The waveform
should show neither an upward
nor
a downward slope anywhere along the
top of the
waveform.
Any
slope
of
the
waveform can
be most easily
seen
with
a sweep speed of approximately 5
µ
seconds
per
centimeter.
Checking
the Bumpiness of the Display
The
next characteristic to look for in the displayed wave
form
is the bumpiness in the top portion of the waveform
immediately
following
the rise.
This is the portion of the
waveform affected
by the
delay line adjustments. To make
this
observation, use
a sweep speed of approximatly .2
µ
sec
onds
per centimeter.
There are
two general types of bumps
to be found in a
poorly
adjusted delay line. The first type is the irregularity
caused
by the
misadjustment of a group of capacitors. This
type
is indicated by a fairly long duration bump. If the
bumps
occur at random intervals along
the delay line, they
are
probably due
to misadjustment of the delay line and can
usually
be
corrected
by a few slight adjustments. However,
I
A)
(A)
6-9
Содержание 535A
Страница 2: ......
Страница 4: ...Type 535A ...
Страница 28: ...Circuit Description Type 535A 4 2 Fig 4 1 Simplified Type 535A Vertical Amplifier ...
Страница 30: ...Circuit Description Type 535A 4 4 Fig 4 2 Simplified Time Base Trigger Circuit ...
Страница 32: ...Circuit Description Type 535A Fig 4 3 Simplified Time Base Generator 4 6 ...
Страница 34: ...Circuit Description Type 535A 4 8 Fig 4 4 Simplified Delay Pickoff Circuit ...
Страница 38: ...Circuit Description Type 535A Fig 4 6 Simplified Alternate Trace Operation Circuitry 4 12 ...
Страница 39: ...Circuit Description Type 535A 4 13 ...
Страница 41: ...Circuit Description Type 535A Fig 4 8 Simplified Low Voltage Power Supply 4 15 ...
Страница 42: ...Circuit Description Type 535A 4 16 Fig 4 9 Simplified Calibrator Circuit ...
Страница 44: ...Circuit Description Type 535A Fig 4 10 Simplified Cathode Ray Tube Circuit 4 18 ...
Страница 126: ...MRU ll IG 61 O Z ...
Страница 138: ...C897 R896 R891 R890 R889 R888 R887 R886 R885 AMPLITUDE CALIBRATOR SWITCH Bottom View 535A A ...
Страница 140: ...C t Aao 3 ELEV 1 TERM 27 TERM 2e TIME BASE B CHA S5 S TYPE 3 5 C S C S C C P E R E A T ER WIRING DIAGRAM ...
Страница 143: ...FIG 1 FRONT SWITCHES TVDC COCA ocril I iCrODC ...
Страница 144: ...70 ...
Страница 145: ...1 FIG 3 SWEEP CHASSIS HIGH VOLTAGE BOARD DELAY SWEEP CHASSIS ...
Страница 146: ...1 FIG 4 POWER CHASSIS RECTIFIER MOUNTING PLATE A ...
Страница 147: ...3 34 79 ...
Страница 148: ...TYPE 535A OSCILLOSCOPE ...
Страница 149: ......
Страница 150: ...FIG 7 CABINET FRAME RAILS ...
Страница 152: ...FIG 8 STANDARD ACCESSORIES 6 TYPE 535A OSCILLOSCOPE ...