![Tektronix 503 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/tektronix/503/503_instruction-manual_1078910042.webp)
Performance Check— Type 503
Fig. 5 -6 . X-Y method o f calcula ting phase difference ( ft) o f two
sine waves.
n.
Repeat the checks as outlined In Table 5-3, and steps
(e) through (k).
TABLE 5 -3
Sensitivity
H o rizo n ta l and
V ertical
G e n e ra to r
Frequency
Maximum
2Trace
S eparation
1 mV/CM
50 kHz
350 kHz
450 kHz
1 mm
2 mV/CM
50 kHz
350 kHz
450 kHz
1 mm
5 mV/CM
50 kHz
350 kHz
450 kHz
1 mm
10 mV/CM
50 kHz
350 kHz
450 kHz
1 mm
20 mV/CM
50 kHz
350 kHz
450 kHz
1 mm
.1 VOLTS/CM
50 kHz
350 kHz
450 kHz
1 mm
.2 VOLTS/CM
50 kHz
350 kHz
450 kHz
1 mm
~AII measurement made w ith a displa y am plitude o f 6 cm, h o ri
zontal and vertical.
2 1
.
Check Common Mode Rejection Ratio (AC)
a. Requirement— Maximum display amplitudes as indi
cated in Table 5-4, with specified signal inputs and deflec
tion sensitivities.
b. Set the HORIZONTAL + and — INPUT switches to
GND.
c. Set the VERTICAL +INPU T to DC, and the -IN P U T
to GND.
d. Set the constant amplitude generator frequency to
50 kHz, and connect the 50-ohm cables to the VERTICAL +
and — INPUT connectors.
e. Set the VERTICAL SENSITIVITY to 1 VOLTS/CM, and
adjust the generator output for a trace 4 cm in amplitude.
f. Change the VERTICAL — INPUT switch to DC, and
change the VERTICAL SENSITIVITY switch to .2 VOLTS/CM.
g. W ith the VERTICAL POSITION control, move the dis
play to the top, then to the bottom of the graticule.
h. Check— Vertical deflection (trace separation) not more
than 2 mm anywhere in the graticule area.
NOTE
In a ll succeeding
Common M ode
Rejection Ratio
checks, the d is p la y should be moved over the
entire g ra tic u le area d u rin g the check, as in d i
cated in steps (g ) and ( h ) .
i. Check the remaining vertical attenuator settings listed
in Table 5-4. The generator output (4 V) is not changed for
the checks in Table 5-4.
j. Set the VERTICAL -+- and — INPUT switches to GND,
and change the HORIZONTAL + and — INPUT switches
to DC.
k. Move the signal cables to the HORIZONTAL INPUT
connectors, and
set the HORIZONTAL SENSITIVITY switch
to .2 VOLTS/CM.
TABLE 5 -4
SENSITIVITY Settings
H O R IZO N TA L/
S ignal
Maximum
VERTICAL
A m p litu d e
D eflection
.2 VOLTS/CM
4 volts
2 mm
.1 VOLTS/CM
4 volts
4 mm
50 m V/CM
4 volts
8 mm
20 mV/CM
4 volts
2 cm
10 m V/CM
4 volts
4 cm
I.
W ith the HORIZONTAL and VERTICAL POSITION
controls, move the display within the graticule viewing area.
m. Check— Horizontal deflection not more than 2 mm.
n. Check the remaining horizontal attenuator settings
listed in Table 5-4, leaving the generator output at 4 V.
o. Change the HORIZONTAL SENSITIVITY switch to 0.1
VOLTS/CM.
p. Change the HORIZONTAL — INPUT switch to GND,
and adjust the generator output for a trace 5 cm long (0.5 V
output).
5 - 9
Содержание 503
Страница 5: ...Fig 1 1 Type 503 Oscilloscope Type 503 ...
Страница 10: ...O perating Instructions Type 503 2 3 Fig 2 2 Type 503 Oscilloscope front panel ...
Страница 23: ...NOTES ...
Страница 33: ...NOTES ...
Страница 49: ...Type 503 colibrotion__ 6 2 ...
Страница 78: ...NOTES ...
Страница 110: ... I N P U T INPUT E X T E R N A L TRIG IN LI N E INPUT INPUT A T Y P E B O 3 O S C I L L O S C O P E ...
Страница 111: ...G Ab A B L O C K D I A G R A M ...
Страница 112: ...T R K a G iE B IN P U T A M P L IF IE R T V p E 5 0 3 O SCILLO SCO PE D ...
Страница 114: ...h S W E E P B A T IN G M U L T I V I 6R A T O R D lS C O N N E D IO D E S EX TYPE 5 0 5 OSC 5 T SCOPE ...
Страница 115: ... D Z 2 I j O H OA uit zul Z Q 3 O 0 7 7 ...
Страница 116: ...c L O a l LLl A uJ u e x a uJ LLl t i ...
Страница 117: ... S W E E P T IM C m 1 1 SW IfeO 5F4 B 3F R I O O V T IM IN G P E S 1 S T 0 P S TYPE 5 0 3 OSC ILLO SCO PE B ...
Страница 118: ...0 S 2 J 7 ...
Страница 119: ... J 0 o lii I in I I I 2 z 3i i pfSg D T IM IN G i S W IT C H ...
Страница 121: ...n I u J a d i ...
Страница 132: ...FIG 1 FRONT ...
Страница 133: ...FIG 1 FRONT TYPE 503 OSCILLOSCOPE ...
Страница 134: ...FIG 2 SWITCHES ...
Страница 135: ...FIG 2 SWITCHES TYPE 503 OSCILLOSCOPE ...
Страница 136: ...FIG 3 CRT SHIELD 2 5 A I ...
Страница 137: ...FIG 3 CRT SHIELD f TYPE 503 OSCILLOSCOPE ...
Страница 138: ...FIG 4 CHASSIS REAI ...
Страница 139: ...TYPE 503 OSCILLOSCOPE ...
Страница 140: ...FIG 5 CABINET 7 ...
Страница 141: ...TYPE 503 OSCILLOSCOPE ...
Страница 147: ...Page 2 of 2 TYPE 505 SCHEMATIC CORRECTION P A R T I A L H O R I Z O N T A L A M P L I F I E R M13 963 568 ...