Performance Check— Type 503
I
v .
V
_
-
n
— /—
r~ r~
r* r* r* r*
\r~ r
r- r
f~ r
r- t
(A ) C 416 misadjusted
(B) C 416 correctly adjusted
Fig. 5 -2 . Typical displays. V ertical A m p lifie r atte n u a to r com pensation check; 7 kHz square-wave inpu t; sweep speed, 2 m illiseconds/cm .
TABLE 5-2
SENSITIVITY
I Std. A m p litu d e
Setting
C a lib ra to r
Deflection
1 mV/CM
1
5 mV
5 cm
~t
1.5 mm
2 mV/CM
|
10 mV
5 cm -+: 1.5 mm
5 mV/CM
20 mV
4 cm ± 1 .2 mm
10 mV/CM
50 mV
5 cm
1.5 mm
20 mV/CM
0.1 V
5 cm ± 1.5 mm
50 mV/CM
0.2 V
4 cm ± 1.2 mm
.1 VOLTS/CM
0.5 V
5 cm -H i .5 mm
.2 VOLTS/CM
1 V
5 cm -+-1.5 mm
.5 VOLTS/CM
I
2 V
4 cm -+-1.2 mm
1 VOLTS/CM
5 V
5 cm -h 1.5 mm
2 VOLTS/CM
10 V
5 cm i t 1.5 mm
5 VOLTS/CM
20 V
4 cm : t 1.2 mm
10 VOLTS/CM 1
50 V
|
5 cm rH 1.5 mm
20 VOLTS/CM
100 V
5 cm - H .5 mm
c. Check that the HORIZONTAL -plNPUT switch is set
to DC, and the — INPUT switch is set to GND.
d. Change the HORIZONTAL DISPLAY switch to HORIZ.
AMPLIFIER (SWEEP DISABLED), and adjust the INTENSITY
as desired.
e. Repeat step 15(e), using the HORIZONTAL SENSITIV
ITY switches. The display w ill be 2 dots.
f. Change the HORIZONTAL + INPUT switch to GND,
and change the HORIZONTAL —INPUT switch to DC.
g. Repeat step 15(e), using the HORIZONTAL SENSITIV
ITY switches. The display w ill be 2 dots.
h. Remove the Standard Amplitude Calibrator signal.
i. Return the HORIZONTAL DISPLAY switch to SWEEP
NORMAL ( X I) .
17. Check Trace Shift Due to Input Grid Current
a. Requirement— Trace shift not more than 1 cm.
b. Set the VERTICAL and HORIZONTAL + and -IN P U T
switches to GND.
c. Set the VERTICAL and HORIZONTAL SENSITIVITY
switches to 1 mV/CM.
d. W ith the short patch cord (item 15 of Recommended
Equipment), ground the VERTICAL +INPU T connector.
e. Change the VERTICAL +IN P U T switch to AC.
f. Check— Vertical shift of the trace not more than 1 cm.
g. Repeat steps (d), (e), and (f) with the VERTICAL — INPUT
connector, and the HORIZONTAL
+
and — INUPUT con
nectors.
h. Remove the patch cord.
18. Check Attenuator Compensation— Vertical
a. Requirement — Waveform with no more than 2%
rounding, overshoot, or tilt.
b. Set the Type 503 controls as follows:
TRIGGER
SLOPE
COUPLING
SOURCE
LEVEL
SWEEP TIME/CM
VARIABLE
HORIZONTAL DISPLAY
AC
1 NT.
FREE RUN
2 mSEC
CALIBRATED
SWEEP NORMAL ( X I )
5-5
Содержание 503
Страница 5: ...Fig 1 1 Type 503 Oscilloscope Type 503 ...
Страница 10: ...O perating Instructions Type 503 2 3 Fig 2 2 Type 503 Oscilloscope front panel ...
Страница 23: ...NOTES ...
Страница 33: ...NOTES ...
Страница 49: ...Type 503 colibrotion__ 6 2 ...
Страница 78: ...NOTES ...
Страница 110: ... I N P U T INPUT E X T E R N A L TRIG IN LI N E INPUT INPUT A T Y P E B O 3 O S C I L L O S C O P E ...
Страница 111: ...G Ab A B L O C K D I A G R A M ...
Страница 112: ...T R K a G iE B IN P U T A M P L IF IE R T V p E 5 0 3 O SCILLO SCO PE D ...
Страница 114: ...h S W E E P B A T IN G M U L T I V I 6R A T O R D lS C O N N E D IO D E S EX TYPE 5 0 5 OSC 5 T SCOPE ...
Страница 115: ... D Z 2 I j O H OA uit zul Z Q 3 O 0 7 7 ...
Страница 116: ...c L O a l LLl A uJ u e x a uJ LLl t i ...
Страница 117: ... S W E E P T IM C m 1 1 SW IfeO 5F4 B 3F R I O O V T IM IN G P E S 1 S T 0 P S TYPE 5 0 3 OSC ILLO SCO PE B ...
Страница 118: ...0 S 2 J 7 ...
Страница 119: ... J 0 o lii I in I I I 2 z 3i i pfSg D T IM IN G i S W IT C H ...
Страница 121: ...n I u J a d i ...
Страница 132: ...FIG 1 FRONT ...
Страница 133: ...FIG 1 FRONT TYPE 503 OSCILLOSCOPE ...
Страница 134: ...FIG 2 SWITCHES ...
Страница 135: ...FIG 2 SWITCHES TYPE 503 OSCILLOSCOPE ...
Страница 136: ...FIG 3 CRT SHIELD 2 5 A I ...
Страница 137: ...FIG 3 CRT SHIELD f TYPE 503 OSCILLOSCOPE ...
Страница 138: ...FIG 4 CHASSIS REAI ...
Страница 139: ...TYPE 503 OSCILLOSCOPE ...
Страница 140: ...FIG 5 CABINET 7 ...
Страница 141: ...TYPE 503 OSCILLOSCOPE ...
Страница 147: ...Page 2 of 2 TYPE 505 SCHEMATIC CORRECTION P A R T I A L H O R I Z O N T A L A M P L I F I E R M13 963 568 ...