Analyze Waveform Data
Description
Window
Hanning
This is a very good window for measuring amplitude accuracy but less so for resolving frequencies.
Use Hanning for measuring sine, periodic, and narrow band random noise. This window works on
transients or bursts where the signal levels before and after the event are signi
fi
cantly different.
Blackman-Harris
:
This is the best window for measuring the amplitude of frequencies but worst at resolving
frequencies.
Use Blackman-Harris for measuring predominantly single frequency waveforms to look for higher
order harmonics.
Using Advanced Math
The advanced math feature lets you create a custom math waveform expression that can incorporate active and reference
waveforms, measurements, and/or numeric constants. To use this feature:
1.
Push
Math
.