Calibration
4-26
1760 Series Component Waveform/Vector Monitor User Manual
Set Horizontal Gain
1.
To set horizontal gain, push the
CONFIG
menu button.
Select
CALIBRATE
, then
CAL SIG ON
and
100 KHZ
.
Select
GAIN CAL ON
.
a.
In
1 LINE SWEEP
(5
µ
s/div), adjust the
HORIZ CAL
control for one-half cycle per major graticule
division; in
2 LINE SWEEP
(10
µ
s/div), adjust for one
full cycle per major division.
b.
Pushing
RESET H CAL
will cancel any gain adjust-
ment and reinstate the factory gain setting.
Set Vector Gain
1.
To adjust vector gain, select
VECTOR
display mode and
push the
CONFIG
menu button. Select
CALIBRATE
, then
CAL SIG ON
and
GAIN CAL ON
.
a.
Adjust
VECTOR CAL
so that the calibration signal
circle falls on the compass rose (outer ring).
b.
Pushing
RESET V CAL
will cancel any gain adjust-
ment and reinstate the factory gain setting.
Содержание 1760 series
Страница 4: ......
Страница 8: ...General Safety Summary 1760 Series Component Waveform Vector Monitor User Manual iv ...
Страница 17: ...Preface 1760 Series Component Waveform Vector Monitor User Manual xiii 1761 Component Waveform Vector Monitor ...
Страница 19: ...Introduction ...
Страница 20: ......
Страница 24: ...Product Description 1 4 1760 Series Component Waveform Vector Monitor User Manual ...
Страница 34: ...Installation 1760 Series Component Waveform Vector Monitor User Manual 1 14 ...
Страница 36: ...Instrument Configuration 1760 Series Component Waveform Vector Monitor User Manual 1 16 ...
Страница 37: ...Getting Started ...
Страница 38: ......
Страница 67: ...Operation Basics ...
Страница 68: ......
Страница 96: ...Using the Menus 1760 Series Component Waveform Vector Monitor User Manual 3 28 ...
Страница 98: ...Remote Operation 1760 Series Component Waveform Vector Monitor User Manual 3 30 ...
Страница 99: ...Making Measurements ...
Страница 100: ......
Страница 124: ...Basic Measurements 1760 Series Component Waveform Vector Monitor User Manual 4 24 ...
Страница 127: ...Appendices ...
Страница 128: ......
Страница 146: ...Appendix A Performance Specification A 18 1760 Series Component Waveform Vector Monitor User Manual ...
Страница 152: ...Appendix B Remote Connectors 1760 Series Component Waveform Vector Monitor User Manual B 6 ...
Страница 160: ...Appendix D Software Version 1760 Series Component Waveform Vector Monitor User Manual D 2 ...
Страница 161: ...Glossary ...
Страница 162: ......
Страница 173: ...Index ...
Страница 174: ......
Страница 180: ...Index 1760 Series Component Waveform Vector Monitor User Manual I 6 ...