Getting Started
1- 4
TDS3000B Series User Manual
Signal Path Compensation (SPC)
The signal path compensation (SPC) routine optimizes the
oscilloscope signal path for maximum measurement accuracy. You
can run the routine anytime but you should always run the routine if
the ambient temperature changes by 10
_
C or more.
To compensate the signal path, do the following steps:
1.
Disconnect any probes or cables from the channel input
connectors.
2.
Push the
UTILITY
button.
3.
Push the
System
screen button to select
Cal
.
4.
Push the
Signal Path
screen button.
5.
Push
OK Compensate Signal Path
. This procedure takes several
minutes to complete.
NOTE
. The signal path compensation does not include calibration to
the probe tip.
Adjusting the Oscilloscope Time and Date
To set your oscilloscope to the current date and time, do the
following steps:
1.
Push the
UTILITY
button.
2.
Push the
System
menu button to select
Config
.
3.
Push the
Set Date & Time
menu button. Use the side menu
buttons to set the date and time values.
4.
Push the
OK Enter Date/Time
menu button to set the instrument
date and time.
Содержание 071-0957-03
Страница 4: ......
Страница 15: ...Getting Started ...
Страница 16: ......
Страница 48: ...Getting Started 1 32 TDS3000B Series User Manual Rear Panel Connectors 1 5 4 2 6 14 2 V DC 400 mA 3 7 8 ...
Страница 51: ...Application Examples ...
Страница 52: ......
Страница 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Страница 81: ...Reference ...
Страница 82: ......
Страница 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Страница 177: ...Appendices ...
Страница 178: ......
Страница 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Страница 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Страница 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Страница 235: ...Glossary ...
Страница 236: ......
Страница 247: ...Index ...
Страница 248: ......