Utility
3- 78
TDS3000B Series User Manual
Factory Cal.
Service personnel use these functions to calibrate the
oscilloscope internal voltage references using external sources. Refer
to your Tektronix field office or representative for assistance with
these processes.
Cal Due Control.
The calibration due notification occurs only in the
power-on screen. Set the controls to
∞
if you do not want to be
notified when calibration is due.
Diagnostic System
Use the System Diag branch to access these functions.
Bottom
Side
Description
System Diags
Execute
Starts diagnostics.
Loop
Once
Executes the diagnostic loop once.
Always
Executes the diagnostic loop continu-
ously.
Until Fail
Executes until a failure occurs.
Error Log
Page Up
Use to see the previous error log
page.
Page Down
Use to see the next error log page.
Key Points
Starting Diagnostics.
To execute the built-in diagnostic routines,
disconnect all cables or probes from the oscilloscope inputs, and then
push the OK Run Test screen button.
UTILITY
Содержание 071-0957-03
Страница 4: ......
Страница 15: ...Getting Started ...
Страница 16: ......
Страница 48: ...Getting Started 1 32 TDS3000B Series User Manual Rear Panel Connectors 1 5 4 2 6 14 2 V DC 400 mA 3 7 8 ...
Страница 51: ...Application Examples ...
Страница 52: ......
Страница 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Страница 81: ...Reference ...
Страница 82: ......
Страница 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Страница 177: ...Appendices ...
Страница 178: ......
Страница 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Страница 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Страница 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Страница 235: ...Glossary ...
Страница 236: ......
Страница 247: ...Index ...
Страница 248: ......