Performance Verification
40
P6467 High-Speed Probe Adapter Instruction Manual
f.
Delete Sections D3 and D2 from the Data group.
g. Delete Sections C3, C1, and C0 from the Control group.
h. Ensure that the Input Radix is set to Hex for the remaining groups.
9. Select the Clock menu and perform the following steps:
a. Change the Clock field to External.
b. Select the rising edge of Clock_2 with no qualification.
10. Select the Trigger menu change it to match Figure 14.
Figure 14: Trigger Menu Setup
11. Repeat steps 7 through 10 for the module_b setups with the following
exceptions:
a. In the Clock menu, select the falling edge of Clock_2 instead of the
rising edge.
b. In the Trigger menu, assert Signal BtoA instead of Signal AtoB, and
change the last event to monitor Signal AtoB instead of Signal BtoA.
Содержание 070-9176-00
Страница 5: ......
Страница 13: ...Service Safety Summary viii P6467 High Speed Probe Adapter Instruction Manual ...
Страница 39: ...Operating Basics 24 P6467 High Speed Probe Adapter Instruction Manual ...
Страница 45: ...Reference 30 P6467 High Speed Probe Adapter Instruction Manual ...
Страница 49: ...Specifications 34 P6467 High Speed Probe Adapter Instruction Manual ...
Страница 51: ......
Страница 65: ...Adjustment Procedures 50 P6467 High Speed Probe Adapter Instruction Manual ...
Страница 76: ...Replacable Parts P6467 High Speed Probe Adapter Instruction Manual 61 1 2 3 4 5 6 7 Figure 2 Exploded view ...
Страница 77: ...Replacable Parts 62 P6467 High Speed Probe Adapter Instruction Manual ...
Страница 81: ...Appendix B P6467 Test Board 66 P6467 High Speed Probe Adapter Instruction Manual ...
Страница 84: ......