Calibration—MicroLab
|
Instruction
Removing The
Keypad/Display
Board
3.
The
MicroLab
|
front
panel
is
held
to
the
chassis
by
six
The
Keypad/Display board
is
located
in
the
top
of
the
screws.
Remove
the
front
panel
screws;
be
careful
that
MicroLab
|
cabinet.
The Keypad/Display
board
is
removed
the
front panel
doesn't
drop
to
the bench surface.
along with
the
MicroLab
|
front
panel
through
the
front
of
NOTE
the
cabinet. Use
the
following
procedure when
removing
thal
Kevosd/Display
board!
Unless
you
are
careful when
removing
the
front
1.
Set
the
cabinet
on
its top,
as
shown
in
Fig.
6-4.
panel and
Keypad/Display
board,
the
key
caps
will
fall
free.
Hold the front
panel
and
circuit
board
together
until
you
can
set
the
Keypad/Display
board
upright
and
remove
the
front
panel.
2.
There are
four
screws
that
mount
the
Keypad/Display
circuit board to
the
MicroLab
|
chassis. Remove
these
four
screws.
Circuit Board
Screws
Front Panel
Screws
2827-24
Fig.
6-4.
The Cabinet
Top
Containing
The
Keypad/Display
Board.
The
Keypad/Display
board
is
removed
through the
MicroLab
|
front panel.