Performance Verification
013-0278-00 Video Display Clamp
19
Test signal (channel 2)
Corrected signal (channel 1)
Figure 16: 60 Hz rejection check (negative)
10.
Set the compliment of the pulse generator to +.
11.
Set the Video Display Clamp
Polarity
to
+
(positive).
12.
Verify that the amplitude of the 60 Hz sinewave component of the waveform
is
≤
70 mV
p-p
(see Figure 17).
Figure 17: 60 Hz rejection check (positive)