7.3. Replacing the particle filter
Replacing the particle filter is necessary when the device control signals the corresponding error. (see
chapter "Description of the control elements").
CAUTION
Whirling up dust is possible.
The particle filter is a disposable filter element. Do not try to clean the filter element.
When replacing the particle filter, the prefilter mat must be replaced as well.
●
Open the service door of the particle filter housing (see chapter 2.1).
●
Lower the lifting device (A) by turning the clamping screw (B). Therefore, use the hexagon key
that is located on the right of the clamping screw.
●
Carefully pull the particle filter (see chapter 2.1) out of the housing.
●
Push the new particle filter back into the particle filter housing.
Only use TEKA spare filters. Otherwise the proper functioning of the unit is
not guaranteed.
●
Elevate the lifting device by turning the clamping screw so that the particle filter is pressed
tightly against the above housing.
●
Close the service door.