Minimum Input Span
The minimum input span is the minimum difference between input 1 and input 2 signals required
to complete the calibration or scaling of the meter. The minimum span is 0.10 mA.
If the minimum span is not maintained, the meter reverts to input 2, allowing the appropriate input
signals to be applied.
Re-Calibrating the Internal Calibration Reference (ICaL)
The Internal Calibration (ICAL) menu, located in the Advanced features menu, is used to recalibrate
the internal calibration reference. Recalibration is recommended at least every twelve months.
Refer to Internal Calibration (ICAL), page 33 for instructions.
4.4.8
Setting the Time Base (
tbAsE
)
The meter calculates total based on rate and a time base of units per second, minute, hour, or
day.
Press the Enter button, at any time, to accept a setting or Menu button to exit without saving
changes. X
4.4.9
Setting the Total Conversion Factor (
totCF
)
Total Conversion Factor is used to convert to a different unit of total display. For example, to
display rate in gallons and total in litters, enter a conversion factor of 3.7854. When rate and total
units are the same, the Conversion Factor should be
1.0000
.
Press the Enter button, at any time, to accept a setting or Menu button to exit without saving
changes.