10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
3-4
OPERATING INSTRUCTIONS
After prolonged operation on battery power, when the battery
voltage approaches the end of its useful range, the "LO BAT"
indicator will be seen on the display. Although the battery cells
are not fully discharged and they will not be damaged by
additional use, this is a warning to the operator that the
battery cells need to be charged.
Use of HTP-100 probes with R1L-BIR1
1.
Connect the BLUE spade lug from the LEFT HTP-100 Probe
to the RED E+ terminal of the R1L-BIR1.
2.
Connect the RED spade lug from the LEFT HTP-100 Probe
to the RED I+ terminal of the R1L-BIR1.
3.
Connect the WHITE spade lug from the RIGHT HTP-100
Probe to the BLACK E- terminal of the R1L-BIR1.
4.
Connect the BLACK spade lug from the RIGHT HTP-100
Probe to the Black I- terminal of the R1L-BIR1.
5.
Hold the RED HTP-100 Probe in your left hand and the
WHITE HTP-100 Probe in your right hand, and push the
probe points against the surface to be measured with
sufficient force to cause the points to retract slightly.