NOTE:
The GEMINI RF Power Meter may be calibrated to either the input
or output port. However, the RF power source must always be connected
to the input port regardless of which port is calibrated. Failure to do so
may result in incorrect readings. Refer to the Calibration Certificate for
port calibration information.
Connect the USB Type B cable to the GEMINI RF Power
Meter 5 VDC power and data transfer port. Connect the
other end of the USB Type B cable to a suitable 5 VDC, 500
mA power source. The GEMINI RF Power Meter display will
illuminate.
NOTE:
The GEMINI RF Power Meter may be connected to any USB power
source capable of supplying 5 VDC, 500 mA. This includes some mobile
device portable power banks and wall chargers. However, for
communication and data collection, connection to a USB 2.0 or higher
port on a computer or other data collection device is required.
Connect the GEMINI RF Power Meter OUTPUT to a suitable
load using torque techniques appropriate for the connector
type.
WARNING:
High RF power levels may be present at the output
of the GEMINI RF Power Meter.
Always connect an appropriate load
to the instrument output.
WARNING:
The GEMINI RF Power Meter should be serviced by
qualified personal only.
Never attempt to open the instrument casing,
disconnect transmission lines, or otherwise service the meter while RF
power is applied to the system.
Failure to comply with these warnings may result in injury or
death to personnel, and damage to equipment.
Connect the GEMINI RF Power Meter INPUT to the RF power
source using torque techniques appropriate for the
connector type.
Allow 10 minutes for the instrument to warm up.