The CEIL function provides 1
° resolution and is used when approaching the target
temperature moving towards zero in either direction, up
or down.
The Trunc function provides 1
° resolution and is used when approaching the target
temperature moving away from zero in either direction, up or down.
See detailed descriptions below.
rndG
1.
AutO: 0.1
° Resolution for readings <
1000
° and 1
° > 999.9
°. AutO is the
default setting.
2.
CEIL: 1
° Resolution with ceiling rounding is used for a decreasing positive
temperature to assure the lower temperature is fully reached or an increasing
negative temperature to assure a higher temperature is fully reached.
For Example: A decreasing (cooling) temperature 20.4
° actual temperature will display
21
° until 20.0
° is reached. An increasing (thawing) temperature
-
20.6
° will display
-21 °
until
-20.0
° is reached.
3.
nOrL: 1
° Resolution with normal rounding, 20.4
° displays 20
°, 20.5
° displays
21 °.
4.
trnC: 1
° Resolution with truncate rounding is used to assure that an increasing
positive temperature is fully reached or a decreasing negative temperature to
assure the lower temperature is fully reached.
For Example: An increasing temperature (heating) 20.6
° will display 20
° until 21.0
° is
reached.
A decreasing temperature (cooling)
-
20.6
° will display
-20
° until
-21.0
° is reached.