Preparation for Use
2-
7
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
4.
If present, carefully remove old batteries being careful to not damage the battery
contacts;
5.
Use only Zinc
-
Manganese Dioxide or Alkaline
-
Manganese Dioxide AA cells that
have been tested by a nationally recognized laboratory to mee
t
the requirements
of UL 60079
-
11, Sixth Edition. Use of any other battery will invalidate the
T
ype
Certificate and create a hazardous condition. The only batteries that TEGAM, Inc.
has secured such testing,
which are approved for use in either a 921B or 922B,
are
listed on the
T
ype Certific
ate
. They are:
•
Energizer E91
•
Duracell MN1500
6.
Observing proper polarity, install three (3) new, AA
alkaline (IEC
LR6, ANSI 15)
batteries;
7.
Re-
install the battery cover
and two (2) retaining screws;
8.
At initial power on after battery replacement, allow approximately 30 seconds for
instrument to stabilize
.
Figure 1: Battery Installation
Making Your First Temperature Measurement
TEGAM’s 900 Series Thermocouple
Thermometers are designed for easy
operation, while still
providing a feature
-
rich experience via the intuitive user interface.
To get started making temperature measurements right away, follow these steps:
1.
Perform
Section 2.7, Battery Installation and Replacement
2.
Connect a compatible temperature probe to the Channel 1
and
/
or Channel 2 input
connector located at the top of the instrument;
To ensure best measurement accuracy, allow several
minutes for the
thermocouple probe and connector to thermally stabilize after
Retaining Screws
Battery
Compartment
Gasket