Tegam 911B Скачать руководство пользователя страница 8

 

 

Preparation for Use 

 

 

2-2 

10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● 

[email protected]

 

 

Safety Notices and Information 

Read this Operation Manual thoroughly before using the instrument to become familiar with its 

operations and capabilities. 

Visually inspect instrument before using. Do not use if unit appears damaged or with any part of the 

case removed.  

 

 

 

 

WARNING

 

 

 

 

MAINTENANCE INSTRUCTIONS WITHIN THIS MANUAL ARE FOR USE BY QUALIFIED SERVICE PERSONNEL 

ONLY. DO NOT ATTEMPT TO SERVICE THIS UNIT UNLESS YOU ARE QUALIFIED TO DO SO.

 

SHOCK HAZARD  

Disconnect all temperature probes and turn the unit off before removing the battery cover.  
Never connect thermocouple leads to any source where more than 42 Volts (peak) could exist 

between the lead and ground.  If it is necessary to make measurements of an  object at elevated 

electrical potential, the user is responsible for obtaining and properly using a probe that provides adequate 

insulation between the surface with elevated potential and the thermocouple wiring. 
Always disconnect probe leads before opening the battery door or the instrument housing. 

Internal circuits can present a shock hazard if leads are connected to a source of elevated potential. 
Do not use this instrument if the housing, probe wiring, probe, or probe handle are damaged or 

distorted. Housings and wire insulation are part of the personnel protection system, and if damaged could 

expose users to elevated potentials.  

 

EXPLOSION HAZARD 

Never use or store this product with batteries installed, or change batteries, in an environment 

where explosive or flammable vapors or dust suspensions may exist.  For thermocouple 

thermometers suitable for use in explosive environments, see TEGAM’s 921A or 922A  Intrinsically Safe 

Thermometers. 
Do not attempt to recharge alkaline batteries. 
Do not put batteries into bags designed to protect parts from electrostatic discharge (ESD).
 

These bags are specially designed with metal shielding which can short circuit a battery. 
Do not expose batteries to extreme heat or fire. Observe all regional laws and regulations when 

disposing batteries. 
Never use this instrument or any temperature probe or sensor inside a microwave oven. 

 

BURN HAZARD 

Do not touch a temperature probe sheath that has been exposed to toxic substances or 

extremely high or low temperatures. 
Do not attempt to measure temperatures beyond the range of the temperature probe.
  Probe 

damage or personal injury could result from exceeding a probe’s maximum temperature rating. 

 

Safety Notices and Information continued on next page . . . . 

 

 

Содержание 911B

Страница 1: ...Y GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 sales tegam com THERMOCOUPLE THERMOMETERS Models 911B 912B Thermocouple Thermometers Operation Manual rev C Manual Part Number 911B 900 Rev C Publishe...

Страница 2: ...or technical data is proprietary information developed exclusively at private expense Computer software and technical data rights granted to the federal government include only those rights customaril...

Страница 3: ...cs 3 4 Auto Power Off 3 5 Backlight and Backlight Timeout 3 6 Hold Function 3 6 Trend Indicators 3 6 Battery Indicator 3 6 Probe Offset 3 7 Open Wire Detection On Off 3 8 Clear Function 3 9 Invalid Me...

Страница 4: ...ators Display Backlight Four 4 LED Backlight with 30 second timeout Display Resolution 0 1 1000 1 1000 Reading Rate 3 Second for Readings and Trend Indicators Battery Type 3 AA IEC LR6 ANSI 15 Alkalin...

Страница 5: ...tude 0 to 4600 m 0 to 15 092 ft Vibration Random 10 500 Hz 0 03 g2 Hz Shock 30g Half Sine Drop 4 Drops from 1 m to Concrete Storage Temp 40 to 71 C 40 to 159 F PHYSICAL CHARACTERISTICS Dimensions 193...

Страница 6: ...6 6100 FAX 440 466 6110 sales tegam com TEGAM Family of Thermometers Thermocouple Thermometers 911B Thermocouple Thermometer Single Input 912B Thermocouple Thermometer Dual Input Data Thermometers 931...

Страница 7: ...y life1 Four 4 digit dual LCD with LED Backlight Four 4 NIST traceable thermocouple types E J K and T Comprehensive real time statistics MIN MAX AVG RNG STDEV and T1 T22 Easy to clean Probe offset fun...

Страница 8: ...ted to a source of elevated potential Do not use this instrument if the housing probe wiring probe or probe handle are damaged or distorted Housings and wire insulation are part of the personnel prote...

Страница 9: ...ross thermocouple leads in excess of normal thermocouple voltage for the selected range Excessive input voltage could result in blown fuse component damage or fire Application of excessive voltage is...

Страница 10: ...batteries CAUTION Always turn the instrument off and disconnect any input connections before replacing the batteries Re install the battery compartment cover before resuming use of the instrument CAU...

Страница 11: ...1 and or Channel 2 input connector located at the top of the instrument To ensure best measurement accuracy allow several minutes for the thermocouple probe and connector to thermally stabilize after...

Страница 12: ...set value See Section 3 10 Probe Offset for more information g Momentarily press to save your selection and move to Channel 2 probe offset if equipped h If desired repeat Step f above for Channel 2 i...

Страница 13: ...er instrument ON or OFF 1 5s Disable auto power OFF while instrument is on 1 5s Enter instrument Setup Menu While in Setup Menu save current value and step to next parameter Toggle display backlight 1...

Страница 14: ...y indicator 1 HOLD function is active 2 T1 and or T2 OFFSET is active1 3 Channel 1 temperature measurement 4 The active thermocouple type 5 Remaining battery life 6 Active temperature unit 7 Channel 2...

Страница 15: ...1 5s to access the Setup Menu The SET annunciator will appear at the bottom of the display and the currently selected thermocouple type will begin to flash From within the Setup Menu press to step th...

Страница 16: ...result is displayed on the second line of the display PARAMETER AVAILABLE VALUES Thermocouple Type E J K T Temperature Units C F K T1 Probe Offset 0 1 increments T2 Probe Offset1 0 1 increments Open...

Страница 17: ...isting statistics data and initiate a new statistics session Press to step through the available statistics Statistics are displayed in the order shown in Figure 7 below For dual channel models the LC...

Страница 18: ...tion Press to hold the current reading and or statistics result and to pause statistics accumulation HOLD is displayed at the top left of the LCD display New measurements are not displayed trend indic...

Страница 19: ...mperature units are changed while a probe offset is active When using a probe offset verify and if necessary correct the programmed probe offset after changing temperature units To set the probe offse...

Страница 20: ...ire Detection allows the unit to detect if a thermocouple probe is connected to the thermometer This feature is not compatible with some thermocouple calibrators and can result in measurement instabil...

Страница 21: ...ications The LCD display indicates when a measurement or statistic is invalid as shown in Figure 9 below INDICATION DESCRIPTION Or The current measurement or statistic is over range for the selected t...

Страница 22: ...temperature to assure a higher temperature is fully reached For Example A decreasing cooling temperature 20 4 actual temperature will display 21 until 20 0 is reached An increasing thawing temperature...

Страница 23: ...hall be acclimated to the controlled environment for a minimum of four 4 hours 3 The equipment listed in Appendix A is required to fully verify the UUT to the expanded instrument uncertainties specifi...

Страница 24: ...able Correction Corrected Reading in Appendix C 16 Compare the Corrected Reading to the tolerances stated in the 2 Sigma Tolerance column of Appendix C to determine whether the UUT measurement is with...

Страница 25: ...el 1 voltage reading 9 Connect the miniature thermocouple connector of the Copper Mini TC Cable to the Channel 1 input of the UUT a For two channel UUTs using the Split Copper Mini TC Cable connect on...

Страница 26: ...ugh 18 for each remaining value in Figure 13 below For single channel UUTs omit Channel 2 values a For two channel UUTs using the single ended Copper Mini TC Cable disconnect the cable from the UUT Ch...

Страница 27: ...27 Press UUT 28 The display will change to rESA showing the actual measured value saved in the previous steps 29 For single channel UUTs skip to Step 35 below 30 For two channel UUTs continue with Ste...

Страница 28: ...ffset is active Set probe offset to correct value for connected temperature probe see Section 3 10 Probe Offset Temperature probe has not stabilized Observe display trend indicators and wait for stabl...

Страница 29: ...during startup If any diagnostic routine detects a malfunction an error will be displayed as shown in Figure 15 below ERROR CODE DESCRIPTION Err ADC Analog to digital converter error Err CJC Cold jun...

Страница 30: ...ion of the problem should be attached to the instrument Many times repair turnaround is unnecessarily delayed due to a lack of repair instructions or a detailed description of the problem This descrip...

Страница 31: ...ssing and repair process RMA Instrument Model Serial Number Company Technical Contact Phone Number Additional Contact Info Service Instructions Evaluation Calibration Only Repair Only Repair Calibrati...

Страница 32: ...oduct is transported to any other locations Repaired products are warranted for the remaining balance of the original warranty or 90 days whichever is greater Warranty Limitations The TEGAM Inc warran...

Страница 33: ...pe The thermocouple cables must be terminated at one end with a male miniature thermocouple connector for connection to the UUT The opposite end should be terminated as appropriate for the thermocoupl...

Страница 34: ...Appendices B i 10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 sales tegam com B EXPANDED INSTRUMENT UNCERTAINTIES Thermocouple Type E...

Страница 35: ...Appendices B ii 10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 sales tegam com Thermocouple Type J...

Страница 36: ...Appendices B iii 10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 sales tegam com Thermocouple Type K...

Страница 37: ...Appendices B iv 10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 sales tegam com Thermocouple Type T...

Страница 38: ...DATA SHEET THERMO COUPLE TYPE STANDARD VALUE C READING C CABLE OFFSET C CORRECTED READING C 2 SIGMA TOLERANCE C E 250 1 30 95 0 29 0 0 23 995 0 28 J 210 0 59 95 0 27 0 0 23 995 0 28 1200 0 66 K 200 0...

Страница 39: ...10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 sales tegam com TEGAM INC 10 TEGAM WAY GENEVA OHIO 44041 CAGE Code 49374 WEB http www tegam com...

Отзывы: