4.2.4.  Manual Zero Air Check 

 
A zero air check can be carried out manually. To do this the TSP inlet needs to be removed and the 
particle filter (supplied with the instrument) needs to be attached to the monitor. Ensure there is a 
good seal around the inlet.  
 
 

 
 
 
 
 

 

The filter will remove 99.99% of particulates from ambient air. Wait 5 minutes and then check the 
readings on the control module. The reported value should be close to 0.0ug/m3.Record the particle 
monitor reading in µg/m

-3

 and the Volt (V) reading from the diagnostics in the maintenance report.  

4.2.5.  Filter Changes 

There are two green filters which are located to the left of the sample filter holder. 
They are designed to provide protection to the pump and optical window. The 
filters will become dirty and must be replaced periodically.  

To do this, unclip the filters from the connectors and replace. Ensure the filters 
are connected the right way in accordance to the flow arrow. These filters can be 
purchased from Aeroqual. Please contact technical support for a quotation. 

 

Note: BQ filter – Sheath air filter (behind) 
            DQ filter – Sample air filter (in front)

 

4.2.6.  Inlet Cleaning 

 
The TSP inlet can be dismantled by unscrewing the 3 screws and separating the head into two parts. 
Use a lint-free cloth wetted with isopropyl alcohol to clean both the inside and outside of the inlet.  

        

 

4.2.7.  Pump Module Removal 

The pump module can be removed from the Dust Profiler by unscrewing the 4 screws on the base 
plate and sliding the module out. Ensure the Dust Profiler is switched off when removing the module. 
The sample, purge and exhaust lines need to be disconnected and all the connected cables removed. 
The two main failures which would require removal of the pump module are: 

1.  A leak is discovered 
2.  The pump requires replacement 

 
In both these cases the pump module will need to be sent back to the Aeroqual factory for repair. 
Please contact technical support for assistance. 
 

Screws 

WWW.DUSTSENTRY.NL  |  +31 70 413 70 50

Содержание Aeroqual Dust Sentry Pro

Страница 1: ...WWW DUSTSENTRY NL 31 70 413 70 50...

Страница 2: ...lation 14 2 1 Assembly of heated inlet 14 2 2 Connect Mains Power 15 3 Dust Sentry Pro Operation and Software 16 3 1 Getting Started 16 3 2 System Checks 18 3 3 Modem configuration 18 3 3 1 Further in...

Страница 3: ...4 Met One MSO 36 9 5 Vaisala Weather Transmitter WXT520 37 9 6 Cirrus MK 427 38 10 Mounting and Site Positioning Guidelines 39 10 1 Mounting 39 10 2 Inlet height 39 10 3 Measurement Interference 40 1...

Страница 4: ...sion 1 Date released April 2014 Description User Guide for Dust Sentry Pro This user guide is a newly created document for the use of the Aeroqual Dust Sentry Pro Date Revision number Description of c...

Страница 5: ...mail notifications weather sensors and GPS The instrument is packaged in a robust lockable enclosure and is light enough for one person site installations The ease of installation as well as the exten...

Страница 6: ...roubleshooting section The optical unit also requires an earth wire to be connected between the housing and the 0VDC line on the power bus Please check that this is fitted if the unit has been replace...

Страница 7: ...particles L 8PC10 Number of particles with diameter larger than 10 m 0 1000000 particles L 1 1 3 Inlet Heater The Dust Sentry Pro uses a 12 V heater on the sample inlet tube to reduce the humidity of...

Страница 8: ...ide range of functions over the USB connection without needing to open the enclosure door such as data logging and various system checks using the supplied Aeroqual Dust Profiler PC software Note 1 Th...

Страница 9: ...other Outputs 4mA 2V 0 of full scale value 12mA 6V 50 of full scale value 20mA 10V 100 of full scale value The module is easily configured via the Operations menu in the PC software and can convert t...

Страница 10: ...on Cirrus MK427 Noise ON OFF ON OFF Cirrus MK427 Noise module only Firmware AUX_MODULE_02 Use for Analogue inputs Met One MSO weather Met One 034b wind Cirrus MK 427 noise 1 2 3 4 Function OFF OFF OFF...

Страница 11: ...1 GND ACTUATOR GND Pin 2 12V LOOP IN Pin 8 LOOP OUT Pin 12 ACTUATOR IN A programming port is also exposed through the side of the module to allow custom programs to be loaded into the module Note Aero...

Страница 12: ...S HSPA 800 850 900 1700 1900 2100MHz Power supply Single supply voltage 6V DC to 55V DC connector 4 pin micro fit 3mm ADC and GPIO inputs Two outputs optocouplers open collector drive 100ma Two inputs...

Страница 13: ...ion needs to be exceeding the trigger point before the relay is activated If Active output on diagnostics is selected the alarm will trigger upon sensor failure or sensor no report If not selected the...

Страница 14: ...t i Connect parts A and B ii Open door of enclosure and remove protective tape from the optical engine iii Insert Inlet Tube Assembly through base mount and fix the three mounting screws Note Ensure t...

Страница 15: ...ctrical regulations 3 Reconnect into the connector ensuring the plugs are fitted securely and the nut gland tightened Plug Nut gland Connector Terminal block Neutral Blue Earth Yellow Green Live Brown...

Страница 16: ...e enclosure 4 Install Aeroqual Dust Monitor software This can also be downloaded from the Aeroqual website at www aeroqual com 5 Launch the Aeroqual Dust Monitor PC software a Select to log in as eith...

Страница 17: ...e Dust Profiler clock with the computer date time 9 Configure the operations a Select Setup Operations b Enter Data Logging Period c If the alarm relay is being used please check Enable Alarm Relay Ou...

Страница 18: ...eing logged properly If there is an issue with one of these checks please refer to Section 4 or contact Aeroqual for technical support 1 With the Dust Profiler on open the door and observe the display...

Страница 19: ...DNS host name is the name you have configured for this instrument Note You must reboot the Dust Profiler after saving the Dyn Org parameters The modem has the following operating modes 1 TCP Client Mo...

Страница 20: ...ent as the host and can maintain bi directional communication using a polling architecture Many enterprise clients turn to mobile virtual network operators MVNO These MVNO s acquire numerous GPRS serv...

Страница 21: ...le interface for communicating with the Dust Profiler monitor via a PC It can be used to configure the instrument initiate the zero calibration routine modify gain factors poll data and display the da...

Страница 22: ...5 PM1 and TSP This is NOT relevant for the Dust Profiler and will read 0 in the event log The feature is for use with a Dust Sentry Update sensor list Update the sensor list on the database Manual mo...

Страница 23: ...gure GPRS Modem Configures the SIM card properties and operating mode of modem Mobile Recipients Configures mobile SMS alerts Email Recipients Configures email alerts Communication Port Allows setup o...

Страница 24: ...message Reset Timer Resets the timer Reset Controller Performs a software reset of the Dust Profiler Control Module 3 4 5 Calibration Control Factors Displays the Dust Profiler s Gain and Offset facto...

Страница 25: ...Firmware version software controlling instrument operation HTR 45 Reserved H3 298 15 Ambient Temperature K TEMA 1 4 Density 1 PM1 measurement TIMA 2 00 Density 2 PM2 5 measurement TEMR 2 00 Density 3...

Страница 26: ...ustment A set flow rate is essential to ensure consistency in the measurement To measure the flow remove the TSP head and connect the volumetric flow meter to the top of the inlet using the adaptor su...

Страница 27: ...ust connection from the base of the enclosure Remove the TSP head and block the PM inlet Connected the pressure end of a flow meter to the exhaust tubing If there is no leak the flow should drop to ze...

Страница 28: ...ected the right way in accordance to the flow arrow These filters can be purchased from Aeroqual Please contact technical support for a quotation Note BQ filter Sheath air filter behind DQ filter Samp...

Страница 29: ...ble quantity of particulate matter sample has been collected 3 Check the sample flows of both instruments to ensure they are within specification 4 Remove the Reference filter and re weigh 5 Download...

Страница 30: ...faulty RF noise in environment Check the electrical connector on the sensor module is firmly connected Check RS485 bus cable continuity Test with a faster computer Try to reduce RF noise Replace bus c...

Страница 31: ...ne Dirty optics Optics need cleaning Send to your authorised distributor or contact Aeroqual technical support Noisy Readings Dirty optics Laser is old Optics need cleaning Send to your authorised dis...

Страница 32: ...ample flow rate 1 0 LPM Alarm Relay Email and SMS optional Display 4 line alpha numeric Data logging 8GB SD card 15 years data storage Averaging period 1 60 minutes Software Data logging and configura...

Страница 33: ...8 Schematic Diagrams 8 1 Pneumatics 8 2 Electrical WWW DUSTSENTRY NL 31 70 413 70 50...

Страница 34: ...8 3 Communication WWW DUSTSENTRY NL 31 70 413 70 50...

Страница 35: ...m s Resolution 0 01 m s Wind Direction Range 0 359 no dead band Accuracy 3 12 m s Resolution 1 Mechanical External construction protection LURAN S KR 2861 1C ASA PC IP65 Size Weight 142 x 160 mm 0 5 K...

Страница 36: ...Mechanical 0 360 Electrical 0 356 Accuracy 4 Resolution 0 5 Mechanical Finish Clear anodized aluminum Sensor Weight 0 81kg Environmental Operating range 30 C to 70 C Wind Speed Range 0 50 m s Accuracy...

Страница 37: ...t resolution accuracy Rainfall duration output resolution Rain intensity range output resolution cumulative accumulation after latest reset 0 01 mm 0 001 inches 5 counts each 10 s increment when dropl...

Страница 38: ...A to IEC 61672 1 2002 Time Weighting Factory set to Fast F IEC 61672 1 2002 Automatic Calibration Electrostatic Actuator System with DC voltage control Measurement Range 30 100dB A Current Loop Output...

Страница 39: ...bracket Note Only one U bolt is used to secure the bottom bracket 5 Add nuts and washers and tighten the bottom U bolt into position 10 2 Inlet height If PM monitoring is related to human exposure th...

Страница 40: ...uipment Demolition construction activities and change to normal transport patterns due to road works etc can significantly affect the data Ensure a record of such events is kept to account for unexpec...

Страница 41: ...b based Browser compatibility Internet Explorer 7 Chrome Safari Alarms Unlimited text SMS email Parameters Up to 20 dynamically configurable Data Warehouse RAID clustered storage multi site Data Forma...

Страница 42: ...www Orbitdata com When the site is first accessed a default landing page will be shown Select Options Login and Settings Enter your User Name and Password Note You can elect to save the login details...

Страница 43: ...h to zoom in on a specific time interval o Hover the mouse over the graph to view the start time and time interval of the graph to be shown if clicked e g 3 hr graph from 3 00pm Mon 16 Sep 2013 o To z...

Страница 44: ...com or SMS 64 21 123 456 Once the alarms have been configured they are visible from the Options Alarm Information view 11 4 3 Downloading Data The data can be downloaded as a CSV file and opened in Ex...

Страница 45: ...ual reserves the right to make changes and improvements to any of the products described in this document without prior notice Under no circumstances shall Aeroqual be responsible for any loss of data...

Страница 46: ...ed service personnel must work on this product ACCESSORIES Use only approved accessories Do not connect incompatible products CONNECTING TO OTHER DEVICES When connecting to any other device read the a...

Отзывы: