7. Check configuration
a. Select Setup
Configuration
b. Check the correct sensors in the Dust Profiler have been configured.
c. These should match the sensors on the invoice and the sensors list in the instrument
logbook.
d. Click Save and Close.
Note 1: The more sensors configured, the larger the data file will be. Therefore, the
download time will be longer if using a remote connection.
Note 2: The Dust Profiler needs to be restarted after a parameter change.
8. Set Dust Profiler Real Time Clock
a. Click on Tools
Update Real Time Clock (this will synchronise the Dust Profiler
clock with the computer date/time)
9. Configure the operations
a. Select Setup
Operations.
b. Enter Data Logging Period
c. If the alarm relay is being used please check
“Enable Alarm Relay Output”
d. Select the output sensor and set the Alarm trigger
point. The alarm will go off when the concentration
reaches above the concentration set for the
selected sensor.
e. Select the reset point of trigger. This is the
percentage of the alarm trigger concentration at
which the alarm will reset.
f.
Select the alarm trigger point average period. This
signifies the time interval at which the average is calculated for control of the alarm.
Note: The alarm average period must be greater than or equal to the data report
rate