HUMIDITY TEMPERATURE METER 

 
 

 

5

VI .Calibration Procedure 

 

Humidity Calibration: 

 

1. Turn the unit off. Press and hold 

MAX/MIN

HOLD

 and 

°C

 keys. While these three keys are 

pressed, turn the power on. Release these keys and all the segments on the LCD will blink. 

2. After step one, within 3 seconds, the user have to press 

°F

 and 

REL

 key at the same time to 

enter the calibration mode or the tester will go back to normal operation mode. 

3. When the tester get into calibration mode, the humidity reading will blink and “CAL1” will appear 

on the second display.   

4. Insert the humidity probe into the standard humidity cavity of 32.8%RH@25°C. Wait until the 

system to stabilize for 20 minutes then press 

MAX/MIN

 button to create the calibration data. If 

the unit recognizes the value is out of tolerance, the unit will sound 2 beeps and still remain at 
"CAL1”mode. If the tester recognizes the value is within tolerance, new calibration data is 
created and the tester will go into “CAL2” mode, which will be indicated by “CAL2” at 2

nd

 display. 

5. Insert the humidity probe into the standard humidity cavity of 75.3%RH@25°C. Wait until the 

system to stabilize for 20 minutes then press 

MAX/MIN

 button to create the calibration data. If 

the unit recognizes the value is out of tolerance, the unit will sound 2 beeps and still remain at 
"CAL2”mode. If the tester recognizes the value is within tolerance, new calibration data will be 
written into the memory and the calibration is done. 

 

Remark: 

1. When the user perform the humidity calibration, the environment should be kept at the stable 

condition (i.e. Constant temperature and constant humidity in the lab.) to in increase the 
accuracy. 

2. After the probe insert into the standard humidity cavity, the operator should wait at least 20 

minutes to let the condition in the cavity to stabilize. 

3. During the calibration, if the user press 

POWER

 button at any time, the tester will go back to 

normal operation mode and no calibration data will be changed. 

4. During the calibration mode, the user can restore the factory default value by press the 

HOLD

 

and 

°C

 buttons at the same time. 

5. Because it takes some time to stabilize the system, we recommend the operator first insert the 

probe into the 32.8%RH@25°C standard cavity and wait for at least 20 minutes, then power the 
unit on and start the calibration process. 

6. During the calibration, all the displayed reading is calibrated with the old calibration data and 

the auto power function is disabled. Until the calibration process is done, the tester will enable 
auto power function again. 

7. During the calibration, the temperature is fixed at °F scale and it is not selectable. 

 

T1 Temperature Calibration: 

 

1. Turn the unit off. Press and hold 

MAX/MIN

HOLD

 and 

°C

 keys. While these three keys are 

pressed, turn the power on. Release these keys and all the segments on the LCD will blink. 

2. After step one, within 3 seconds, the user have to press 

°F

 and 

HOLD

 key at the same time to 

enter the calibration mode or the tester will go back to normal operation mode. 

3. When the tester get into calibration mode, the temperature reading will blink and “CAL1” will 

appear on the second display.   

TECPEL

Содержание DTM-322

Страница 1: ...REL HOLD C F T1 K TYPE T2 POWER UP OPTIONS C F 0 TEMP 20 C HUMIDITY 328 200 4 F 60 C 2498 1370 RH 140 F 100 F C DTM 322 TECPEL ...

Страница 2: ...311 01 JUL 2000 TECPEL ...

Страница 3: ... Instructions 3 4 1 Power Up 3 4 2 Humidity and Temperature Measurement 3 4 3 Connection the Thermocouples T2 channel 3 4 4 Selecting the Temperature Scale 3 4 5 Data Hold Operation 3 4 6 MAX MIN Operation 4 4 7 Relative Operation 4 4 8 Auto Power Off 4 4 9 Low Battery Condition 4 4 10 Digital Output 4 VI Calibration Procedure 5 TECPEL ...

Страница 4: ...r sensor and K type thermocouple This operations manual contains general product information and specification III Specifications Numerical Display 4 digital Liquid Crystal Display Measurement Range Humidity 0 100 RH Temperature T1 20 60 4 140 T2 200 1370 328 2498 Resolution Humidity 0 1 RH Temperature T1 0 1 0 1 T2 200 200 0 1 200 1370 1 200 200 0 1 else 1 Accuracy Humidity 2 5 RH at 25 Temperatu...

Страница 5: ...on Meter 186 L 64 W 30 H mm 10 8 L 2 5 W 1 2 H inch Probe 190 L 15 D mm 7 5 L 0 6 D inch Accessories Instruction Manual 9V Battery Carrying Case Probe Holder K type bead thermocouple probe IV Symbol Definition and Button Location Centigrade and Fahrenheit indication RH Relative Humidity indication MAX The Maximum value is now being displayed MIN The Minimum value is now being displayed This indica...

Страница 6: ... Fahrenheit F by pressing F button and vice versa to Celsius by pressing C button 4 5 Data Hold Operation The user may hold the present reading and keep it on the display by pressing the HOLD button When the held data is no longer needed one may release the data hold operation by pressing HOLD button again When the meter is under Data Hold operation the MAX MIN and C F UREL button are disabled whe...

Страница 7: ...y default when the meter is powered on it is under auto power off mode The meter will power itself off after 30 minutes if no key operation and no RS232 communication Combination at power on can disable auto power off One may press and hold HOLD button and then power on the meter and there will be two successive beeps to indicate that auto power off is disabled and the will not show up 4 9 Low Bat...

Страница 8: ...he user perform the humidity calibration the environment should be kept at the stable condition i e Constant temperature and constant humidity in the lab to in increase the accuracy 2 After the probe insert into the standard humidity cavity the operator should wait at least 20 minutes to let the condition in the cavity to stabilize 3 During the calibration if the user press POWER button at any tim...

Страница 9: ... mode Remarks 1 After the probe is in the calibration cavity wait at least 20 minutes to stabilize the system 2 During the calibration the user can leave the process by pressing the POWER button at any time and the calibration data will be kept unchanged 3 During the calibration mode the user can restore the factory default value by press the HOLD and C buttons at the same time 4 Because it take s...

Страница 10: ...www tecpel com tw www tecpel com TEL 886 2 2218 3111 FAX 886 2 2218 3222 Email sales tecpel com tecpelco gmail com TECPEL ...

Отзывы: