19
ENGLISH
07
INSTALLATION
PICT.22
EUROPA B EVO
EUROPA B EVO 24
A
610 mm
B
590 mm
730 mm
C
50 mm
D
410 mm
E
426 mm (front)
455 mm (rear)
410 mm (front)
444 mm (rear)
F
400 mm
515 mm
1 - Install the autoclave in an environment suitable for carrying out sterilization procedures.
2 - The room must be adequately lit and ventilated, as provided for in the directives in
force
3 -
Install the autoclave away from heat sources and water splashes.
4 - Position the autoclave on a surface that is able to bear its weight. The minimum
recommended load bearing capacity of the surface is 80 kg.
5 - The supporting surface must be perfectly horizontal and must not have inclinations.
6 - Place the autoclave at a height that allows the user to inspect and easily clean all parts
of the sterilization chamber.
7 - Open the door of the autoclave and remove from inside the sterilization chamber all
packages which contain the single accessories.