NOTES:

-  It  is  recommended  to  turn  off  the  measurement  mode  by  pressing  the  button 

     

before moving the sensor, in order to save the receiver settings and eliminate unpleasant 
noise in the headphones.

- During leak detection it is not recommended to move the sensor and use the memory 

functionary earlier than 10 seconds after placing the sensor on the ground and turning on 
the measurement mode

- Do not change control settings as you move along the route, in order to save the relative 

value of signal level.

In order to enter the review mode: 

1.

 Turn off measurement 

mode by pressing  

2.

 Press the memory 

button  

3.

 Review memorized fields 

using the

◄/► 

buttons

In order to leave memory mode press button  

 - you enter the launch window, and then to 

return to the measurement mode press «start» 

 

When the receiver is powered off, the recorded data is not saved!

3.3  Conduct fine tuning of receiver filter

General principle of the filter setting is the gradual narrowing of band pass in order to isolate the 

sound of leakage and maximum suppression of all other sounds.

1.

 Place the acoustic sensor 

over  the  suspected  leak  point 
and commence filter setup.

2.

  On  hearing  a  sound  like 

the sound of fluid leakage turn 
on  the  band  pass  filter  by  the 
button 

 

3.

 

Switch to Spectrum 

window by pressing visual 
indication type button 
twice

  

  

Suspected 

leak point

 Leak detection АP-027

(10-06-15)

Содержание Success AT-407N

Страница 1: ......

Страница 2: ...f operation Water leak detector Success 407N is designed for water leak detection from metal or plastic pipelines at depths of up to 6 m underground water leak detection from the pipelines inside the...

Страница 3: ...5 filter button broadband on off 12 headphones jack 6 memory button 13 sensors connector 7 start pause button measurement mode 14 protective insert Technical specifications of the Receiver are listed...

Страница 4: ...axis Check the charge status of the batteries the charge should be no less than 4 0 V If the battery charge is low replace the batteries Check the accuracy of sensor connection If a no sensor symbol i...

Страница 5: ...in the spots where the signal level is the highest 7 We recommend you to record the points where signal level is the highest into the memory of the device by pressing the memory button 7 1 Browse the...

Страница 6: ...er to enter the review mode 1 Turn off measurement mode by pressing 2 Press the memory button 3 Review memorized fields using the buttons In order to leave memory mode press button you enter the launc...

Страница 7: ...alid signal and lowest number of light stripes interference signifies correct choice of filter 10 Switch to Scale mode by pressing visual indication button Without changing the settings examine the su...

Страница 8: ...ected modes and parameters animated scale for input signal level numeric value and animated scale of output signal level graph moving diagram of the output signal level output signal frequency spectru...

Страница 9: ...ed battery and the device will automatically switch off after type of the sensor connected Sensor not connected Acoustic sensor AD Electromagnetic sensor EMD type of signal received when working with...

Страница 10: ...output overload Filtering Displays the frequency range scales with numerical and graphical representation of the pass band During work with AD pass band filter broadband DuringworkwithEMD central freq...

Страница 11: ...lue of the signal in the mode current value of the signal in the mode and maximum current value of the signal for the previous second pulse amplitude 5 Acoustic signal spectrum window It displays the...

Страница 12: ...can be called for by pressing the button The display shows the industrial frequency spectrum of 0 40 8 60 kHz Two segment columns display the current and minimum values of the frequency components of...

Страница 13: ...y natural sound is used When working with EMD in the mode natural sound to the phones the adopted highactive frequency8192Hzand33kHz beforeplaying areconvertedintoawell acceptable low frequencies of 8...

Отзывы: