Installation
Proline Promass F 500 EtherNet/IP
30
6.1.3
Special mounting instructions
Rupture disk
Information that is relevant to the process: → 271.
L
WARNING
Limited functional reliability of the rupture disk.
Danger to persons from escaping fluids!
‣
Do not remove the rupture disk.
‣
When using a rupture disk, do not use a heating jacket.
‣
Make sure that the function and operation of the rupture disk is not impeded through
the installation of the device.
‣
Take precautions to prevent damage and danger to persons if the rupture disk is
actuated.
‣
Observe information on the rupture disk sticker.
The position of the rupture disk is indicated on a sticker beside it.
The transportation guard must be removed.
For information on the dimensions: see the "Mechanical construction" section
The existing connecting nozzles are not intended for the purpose of rinsing or pressure
monitoring, but instead serve as the mounting location for the rupture disk.
In the event of a failure of the rupture disk, a discharge device can be screwed onto the
internal thread of the rupture disk in order to drain off any escaping medium.
RUPTURE DISK
i
2
3
1
DN 8 ( ")...150 (6")
!
⁄#
DN 250 (10")
2
3
A0028903
1
Rupture disk label
2
Rupture disk with 1/2" NPT internal thread with 1" width across flat
3
Transport protection
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology.
Calibration takes place under reference conditions→ 261. Therefore, a zero point
adjustment in the field is generally not required.