Proline Promass F 500 EtherNet/IP
Technical data
271
in which the process pressure is greater than 2/3 of the secondary containment burst
pressure.
For information on the dimensions: see the "Mechanical construction" section of the
"Technical Information" document
Rupture disk
To increase the level of safety, a device version with a rupture disk with a trigger pressure
of 10 to 15 bar (145 to 217.5 psi)can be used (order code for "Sensor option", option
CA
"rupture disk").
Rupture disks cannot be combined with the separately available heating jacket .
Special mounting instructions: → 30
For information on the dimensions: see the "Mechanical construction" section
Flow limit
Select the nominal diameter by optimizing between the required flow range and
permissible pressure loss.
For an overview of the full scale values for the measuring range, see the "Measuring
range" section → 252
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained
solids): flow velocity < 1 m/s (< 3 ft/s).
• For gas measurement the following rules apply:
– The flow velocity in the measuring tubes should not exceed half the sound velocity
(0.5 Mach).
– The maximum mass flow depends on the density of the gas: formula → 252
To calculate the flow limit, use the
Applicator
Pressure loss
To calculate the pressure loss, use the
Applicator
Promass F with reduced pressure loss: order code for "Sensor option", option
CE
"reduced
pressure loss"
System pressure
16.10 Mechanical construction
Design, dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section.
Weight
All values (weight exclusive of packaging material) refer to devices with EN/DIN
PN 40 flanges.
Different values due to different transmitter versions: