Installation and Setup

 

A sketch or layout diagram of the system may help you when installing the 
components. The component locations will be determined by the tank inlet hoses, 
the tank switcher outlet connection, and the distance to the control box. The Tank 
Switcher Manifold Assembly also needs a mounting structure and a place to 
mount the control box. 
 
1.  For most systems the tank switcher 

manifold assembly should be mounted 
using channel steel and clamps around 
the vacuum jacketed pipe section. The 
tank switcher mounting needs to be 
capable of holding a minimum of 150 lbs
(70 kg). 13/16” channel steel strut should 
be large enough to support the manifold. 

 

The photo shows an example of a well mounted tank assembly. Be sure to 
mount the manifold in three places (shown with arrows on in the photo), one 
clamp on each of the three ‘legs’. 
 

2.  The Control Box can be located up to 5 

meters (16 feet) from the Tank Switcher 
Manifold with the standard cable. Other 
length cables are available from Technifab. 
Refer to the specifications section for the 
control box electrical and temperature 
requirements. 

 
 
3.  Mount the control box using the holes, four total, at the 

top and bottom of the control box. Connect the control 
cable to the bottom of the control box and to the junction 
box on the manifold assembly, using proper cable 
routing/mounting as needed. The control box comes with 
a standard NEMA 5-15R male power connector at the 
end of the cable. Do not plug the unit in at this time. 

 

Содержание Cryogenic Tank Switcher

Страница 1: ...LT 33020 2 2013 Technifab Products Inc 1 Technifab Vacuum Jacketed Tank Switcher Installation Manual...

Страница 2: ...013 Symbols Used in this Manual Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury Caution indicates a potentially hazardous situation whi...

Страница 3: ...chnifab Products Inc recommends routing the discharge lines from pressure relief devices to a safe place of discharge The outlet of the discharge line should not be in a location where it could injure...

Страница 4: ...her Manifold Assembly with Junction Box Note Most tank switchers include a cryogenic inlet filter installed inside the inlet female bayonets You should use an orange colored bayonet clamp two shown to...

Страница 5: ...ld be large enough to support the manifold The photo shows an example of a well mounted tank assembly Be sure to mount the manifold in three places shown with arrows on in the photo one clamp on each...

Страница 6: ...cables are connected on the right side from the bottom up 6 Hookup the inlet and outlet piping to the manifold If the tank switcher inlets have inline cryogenic filters be sure to use the supplied ora...

Страница 7: ...econfigured at the factory For shop air the Pneumatic Actuated Valve Guidelines Technifab part number literature LT 35569 has a more complete description of the differences and shop air requirements F...

Страница 8: ...deliver liquid nitrogen Press the onscreen Run Stop button press the onscreen reset button and press the Run Stop button again A longer Switcher Timer SP setting may be necessary Error message on scre...

Страница 9: ...tts 110 VAC nominal 1 7 amp maximum Alarm Output 24 VDC 0 2 amp max current Product Specifications Technifab Cryogenic Tank Switcher Support Information Technifab continuously seeks to improve its pro...

Отзывы: