ADDITIONAL INFORMATION

User maintenance

1.

Cleaning

Your Sample Concentrator can be cleaned by wiping with a cloth dipped in soapy water. Ethanol and
formaldehyde may also be used. Do not use abrasive cleaners.

2.

Carriage friction

Height adjustment is by means of a friction drive. To allow for wear the friction pressure may be
increased by tightening the 4 screws on the rear of the carriage sufficiently to give adequate grip.
Ensure that all four are tightened equally.

3.

To replace the sealing pad

After many repeated needle insertions the self sealing property of the sealing pad may start to
degrade. Spare pads can be obtained from your supplier or from Techne. Replacing the pad is simply
accomplished as follows:

Remove the gas chamber from the carriage, disconnect the gas pipe and place upside down on a
suitable work surface.

Remove the four nuts securing the matrix plate.

Lift off the matrix plate leaving the four spacers in position.

Lift off the sealing pad. Ensure that the support strip underneath the sealing pad and the four spacers
remain in place.

Replace the sealing pad with the new one. Reassembly is the reverse of the above procedure.

For any other servicing requirements please contact your supplier or Techne’s service department.

Содержание FSC496D

Страница 1: ...Sample Concentrator OPERATOR S MANUAL Issue 10 8 08...

Страница 2: ......

Страница 3: ...fety 3 Specification 3 Installation 3 Diagram of Sample Concentrator 4 Guarantee 5 OPERATION 6 Preparation 7 Operation 8 ADDITIONAL INFORMATION 9 User maintenance 9 Solvent evaporation temperatures 10...

Страница 4: ...d over the surface of the samples displaces the evaporated solvent from above the liquid s surface This significantly increases the rate of sample concentration The unit provides accurate and reproduc...

Страница 5: ...Blocks Gas Intake Nozzle Diameter 6 35 mm 1 4 Height 530 mm Width 240 mm Depth 295 mm Weight 3 5 Kg INSTALLATION For packing purposes the Sample Concentrator is supplied unassembled in three parts the...

Страница 6: ...f 2 psi will cause gas to be vented and wasted 4 Place the Techne Dri Block heater onto the base of the Sample Concentrator It is positioned by means of two spigots on the base board which align with...

Страница 7: ...under this guarantee should be obtained from the supplier Notwithstanding the description and specification s of the units contained in the user manual Techne hereby reserves the right to make such ch...

Страница 8: ...ng with the type of insert block used in the heater These are listed in the following table As an example if you use the block for twelve 15mm test tubes block number F3506 you would place a needle in...

Страница 9: ...e removed and replaced many times as the pad is self sealing INSERTING THE NEEDLES 4 Replace the gas chamber onto the carriage and tighten the locking knob Re connect the gas pipe if necessary C C C A...

Страница 10: ...amber so that the tips of the needles enter the test tubes Position them at the required height above the liquid surface 3 Set the heater to the required temperature and adjust the gas flow to the des...

Страница 11: ...le insertions the self sealing property of the sealing pad may start to degrade Spare pads can be obtained from your supplier or from Techne Replacing the pad is simply accomplished as follows Remove...

Страница 12: ...35 Atmospheric 1 2 dichloroethane Ethylene chloride 83 210 1 2 dichloroethene Cis acetylene dichloride 60 479 1 2 dichloroethene Trans acetylene dichloride 48 751 2 propan 2 yloxypropane Isopropyl eth...

Страница 13: ...ethane Carbon tetrachloride 76 300 1 1 2 2 tetrachloroethene Tetrachloroethylene 121 53 Oxolane Tetrahydrofuran THF 67 357 Toluene Methylbenzene 111 77 1 1 2 trichloroethene Trichloroethylene 87 183 O...

Страница 14: ...B B B B B B B B B B B B B A A A A A A A A A A A A A A A A A A A A A A A A A A D D D D D D A A A A A A A A A A A A A A A C C C C C C C C C C C C C C C C C F E E E E E E F F F F F B B B B B B B B B B B...

Страница 15: ...needles 100 x 127mm long 6101609 Base board 6101608 Spigot 6101604 Sealing pad 6101606 Spacer 6101605 Support strip Spare and alternative heads complete 6010080 Standard three block matrix head 61042...

Страница 16: ...x France Tel 33 1 64 45 13 13 Fax 33 1 64 45 13 00 e mail bsf bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano I...