Calibration Of The Thermoregulator
Remember that if you change the calibration from that set at the factory you may change the calibration
at all temperatures. You may get different calibration with different baths and/or liquids.
In order to ensure that the calibration you are setting is correct, you will need to use an independent calibrated
probe or thermometer.
Set the temperature display to the particular temperature at which you require to control. Measure the actual tem-
perature of the bath liquid using a calibrated probe or thermometer.
If the calibration is not correct then you can follow this procedure.
1. Hold down the Up and Down buttons and then press the Set button at the same time for 5 seconds.
2. The display will change from the bath temperature to “EEEE”.
3. Press the ‘SET’ button and either the ‘UP’ or the ‘DOWN’ button to adjust the display to the same temperature
as the measured value.
4. Press ’UP’ and ‘DOWN together to confirm the value. The display will return to the bath temperature and the
unit will control with the new calibration parameters.
Servicing And Repair
NOTE THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL.
REMOVING THE TOP CASE EXPOSES POTENTIALLY LETHAL MAINS VOLTAGE
.
This product range does not require any routine servicing.
NOTE:
There are no internal user replaceable parts.
In the event of product failure, it is recommended that any repair is only undertaken by suitably
qualified personnel. For advice, please contact Techne quoting the model and serial number.
Only spare parts supplied by the manufacturer or its agent should be used. Fitting of non-approved
parts may affect the performance of the safety features of the instrument.
If in doubt, please contact Techne.