Manual TECHKON DENS

How to measure

After switching DENS on, it is instantly ready to measure. Just position the measure-
ment aperture on top of the measurement sample and press shortly with the index 
finger the DENS button. Within less than one second the resulting measurement value 
will appear on the display of the device.

Please ensure that the device always has a firm stand on a flat surface. There must not 
be a distance between the measurement head and the sample where light could pass 
through, because this can influence the measurement. 

The color of the background material underneath the measurement sample can have 
an influence on the measurement result. Different technical standards describe which 
backing material to use. In the printing industry the following procedure is widely used: 

  white backing for proofs and

  black backing for prints. 

Calibration

Prior to every series of measurements a calibration should be carried out. After a cali-
bration the device is long term stable, however it is recommended to perform a cali-
bration before starting a new measurement task just to be sure the device is working 
properly.

For density measurements according to Status ISO E / Pol the calibration is carried out 
on paper white. This will ensure that the density values are in direct ratio to the ink 
thickness which is applied on the paper. When there is a change in the paper type, a 
new calibration has to be done. 
When a calibration according to Status ISO T / no Pol is carried out an absolute white 
standard is used (see enclosed reference card).
For dot % measurements in transmission a calibration is done on a unexposed spot of 
the film to be sure, that no “veil” on the film will falsify the dot % value.

A calibration is as easy to carry out as a measurement: Place the measurement head 
on top of a blank area of the paper sheet or the absolute white standard. The calibra-
tion can be started now by pressing the CAL button. A successful calibration will be 
indicated be three horizontal lines. The device is now ready for measurements.
During the calibration process always take care that the instrument is on a flat, stable 
surface and that the material backing the paper is the same as for the following mea-
surements (white, gray or black). When the paper sample is printed on both sides, it 
is recommended to use a black backing material in order to avoid mismeasurements 
due to color shining through the back of the opaque paper.

Содержание DENS

Страница 1: ...Manual Color Densitometer DENS...

Страница 2: ...can not fully guarantee the complete absence of errors TECHKON GmbH and the authors can not be held liable neither for any possibly incorrect statements nor their consequences Products that are not m...

Страница 3: ...5 minutes you will know how to work with DENS The second chapter will show you detailed insight into the measurement functions Please get the device registered by using the registration card which you...

Страница 4: ...orking with DENS 9 1 5 Care and maintenance 11 ChaptEr 2 Description of the measurement functions 2 1 Measurement function D 14 2 2 Measurement function D 15 2 3 Measurement function T 16 2 4 Measurem...

Страница 5: ...amental densitometric quality control on the basis of the reliable high accuracy well known from all TECHKON instruments The DENS indicates densities dot area and gray balance clearly legible on a lar...

Страница 6: ...user friendly measurement button Durable precision Handy and compact low weight Well designed Two different versions We supply TECHKON DENS in two device versions DENS Status ISO E Pol Density filter...

Страница 7: ...manual A comprehensive DENS manual can be downloaded under www techkon com index php dens html Optional accessories Print Control Strip TECHKON TCS the Digital Print Control Strips TECHKON TCS can be...

Страница 8: ...r at the side of the device is part of the package K C M Y 80 40 0 K C M Y K C M Y 80 40 0 K C K C K C M Y K C M Y K C M Y BAL BAL C M BAL BAL C Y M Y BAL M Y BAL 0 BAL K 5 4 3 2 80 40 80 40 80 40 80...

Страница 9: ...ttery status is displayed as a battery symbol with three segments in the lower right corner of the display when the device is turned on When the battery turns empty the last remaining segment of the b...

Страница 10: ...working properly For density measurements according to Status ISO E Pol the calibration is carried out on paper white This will ensure that the density values are in direct ratio to the ink thickness...

Страница 11: ...y labels at the bottom of the device This could lead to faulty measurements because the direct contact which is important for the correct optical field of depth might not be maintained Warranty The wa...

Страница 12: ...evice case 4 When reassembling do not tighten the screws too much otherwise they might break Replacement of the rechargeable battery The rechargeable Li Ion battery has a high lifespan However after a...

Страница 13: ...ch of the reference and press the CAL but ton 3 Place the measurement head on the Cyan solid density patch of the reference and press the DENS button 4 The desired value can be set respectively by pre...

Страница 14: ...tween 1 00 and 2 00 Chapter 2 Description of the measurement functions The current measurement function is indicated on the right side of the display A switch to the next mode is done by pressing the...

Страница 15: ...posed of different portions of Cyan Magenta and Yellow It does not include black ink 1 Measurement 2 Measurement 2 2 Measurement function D Dot area measurement in reflection For the function Dot Area...

Страница 16: ...he luminosity of the light table Av erage bright light tables allow you to measure densities up to D 4 00 and over Please note The emulsion side of the film has to be always on the upper side 2 4 Meas...

Страница 17: ...nt time 0 5 seconds per measurement White reference Relative ISO E or absolute ISO T Density filter ISO E with polarization filter optional ISO T without polarization filter Density measurement range...

Страница 18: ...s maintenance free The measurement aperture has to be kept clean from dust It can be cleaned with clean compressed air and an optics brush We recommend a functionality check up every 24 months at the...

Страница 19: ...al standards DIN 5033 part 1 9 CIE and ISO 5 3 The accuracy of the device is checked during the manufactur ing by spectral measurements on color samples which were measured and certified by the German...

Страница 20: ...anforderungen gem eu richtlinie 2004 108 ec entspricht und dass die folgenden Normen angewandt wurden complies with the essential requirements of the 2004 108 EC Directive if used for its intended us...

Страница 21: ...Please put my E mail address on the mailing list for the TECHKON Newsletter TECHKON GmbH Wiesbadener Str 27 D 61462 K nigstein Via telefax to 49 0 6174 9244 99 Please send this registration card by ma...

Страница 22: ...TECHKON GmbH Wiesbadener Str 27 D 61462 K nigstein T 49 0 6174 924450 F 49 0 6174 924499 info techkon com www techkon com Erfolg ist messbar...

Отзывы: