20
Example 1
If the activation point is programmed to 100 cm and the deactivation point is programmed to
90 cm, when the level is zero the output will be deactivated. When the level reaches a value
of 100 cm the output will be activated until the level falls below 90 cm.
Example 2
If we program an activation point of 90 cm and a deactivation point of 100 cm, when the
level is zero the output will be activated. When level reaches a value of 100 cm the output will
be deactivated until the level falls below 90 cm.
7.5.1.2 No action
The relay will be always at rest.
7.5.2 Analog
output
The level value corresponding to 4 mA and 20 mA can be programmed and the calibration
for these two values can be made as well.
Содержание MT03L Series
Страница 30: ...30 15 DIMENSIONS Panel hole...