5. Maintenance
& Cleaning
42
Instructions for Use for SUNRISE No. 30086638 Rev. No. 2.9
2021-07
5. Carefully push the filter and retaining ring into the filter carriage, using a
wooden or rubber rod.
5.2.2
Gradient Filter Carriages for Tuneable Wavelength Option
Note
The instrument specifications can only be guaranteed if genuine Tecan
parts are used.
Using the option tuneable wavelength selection, the instrument is fitted with a
special gradient filter, which allows the selection of any wavelength of light
between 400 and 700 nm. For measurements in the ranges of 340 - 399 nm and
700 - 750 nm the appropriate 4filter carriage, containing the required wavelength
filters, must be inserted into the instrument.
The filter carriage can be exchanged for another carriage, which contains other
filters; the instrument can store the data for up to eight filter carriages.
To change the filter carriage, please follow the procedure outlined in the
appropriate software Instructions for Use.
Note
The SUNRISE specifications are different if the instrument is equipped with
the 4filter option, 6filter option or tuneable wavelength option.
Validation of the system has to be performed with those filters, which
finally are actually used during measurement.
Specifications for SUNRISE with 4Filter Option / 6Filter Option
PARAMETERS
CHARACTERISTICS
Accuracy:
(492 nm)
0.000 - 2.000 OD
(492 nm)
2.000 - 3.000 OD
< (1.0 % + 0.010 OD) *
< (1.5 % + 0.010 OD) *
Precision:
(492 nm)
0.000 - 2.000 OD
(492 nm)
2.000 - 3.000 OD
< (0.5 % + 0.005 OD) *
< (1.0 % + 0.005 OD) *
Linearity:
(400-750 nm)
0.000 - 2.000 OD
(400-750 nm)
2.000 - 3.000 OD
< 1 %
< 1.5 %
(400 - 750 nm) 0.000 - 2.000 OD
(400 - 750 nm) 2.000 - 3.000 OD
R² >= 0.999
R² >= 0.999
Specifications for SUNRISE with Gradient Filter of Tuneable Wavelength
Option
PARAMETERS
CHARACTERISTICS
Accuracy:
(492 nm)
0.000 - 2.000 OD
< (1.5 % + 0.010 OD) *
Precision:
(492 nm)
0.000 - 2.500 OD
< (1.0 % + 0.005 OD) *
Linearity:
(492 nm)
0.000 - 2.500 OD
< 2 %, R² >= 0.999
* better than x % of measurement value plus corresponding OD value
Note: All deviations from the measurement value are meant in positive and negative direction.