17
a
Memory number
The currently selected memory number is shown.
b
HOLD indicator
When the hold function is active, a white indicator
appears beneath the word HOLD. When the zone defini-
tion hold function is active, a gray indicator appears.
c
Hold function setting
The currently selected hold mode is shown.
d
JUDGE indicator
When judgment output is active, this indicator appears
white.
When Judge Signal is enabled, this appears white if
the JUDGE input terminal is ON and black if it is OFF.
When Judge Signal is disabled, it is always white.
e
Judgment type
HH/HI
These light to show judgment output operation for
indicator values greater than the setting values for the
high high and high limits.
OK
This lights to show OK judgment output operation
when the indicator value is equal to or higher than the
low limit and equal to or less than the high limit.
LL/LO
These show judgment output operation for indicator
values lower than the setting values for the low low
and low limits.
NOTE
Judgment output operation changes according to the
Comp. Pattern, Comp. Output Pattern and Hysteresis
settings.
The above example is of operation when the Comp.
Pattern setting is LL/LO/OK/HI/HH, the Comp. Output
Pattern is “Standard Output” and the Hysteresis setting
is 0.
f
Indicator value
1. Names and functions of parts
g
High limit, low limit, high high limit and
low low limit setting values
Each setting value is shown.
If high high limit and low low limit are disabled, the LOCK
settings are shown.
h
Bar meter
This is an analog representation of the indicator value.
The highest value of the bar meter range is set with the
Max. Disp. Value setting.
If the low low limit and low limit are zero or higher, the
bar meter will show from zero to the maximum display
value.
If the low low limit or low limit is less than zero, the bar
meter will show zero as the center value and show a
range from the maximum display value to negative that
value.
NOTE
If the bar meter zero position is at the center, a white line
will appear at the zero position.
i
Unit
j
LOCK settings
This shows the lock status.
CAL: Cal. Value Lock
SET: Setting Value Lock
KEY: Key lock
These do not appear when the high high limit and low
low limit are enabled.
NOTE
There are two key lock types.
Press and hold ESC + ENT
This is indicated on the front panel.
This cannot be used to unlock when key lock has been
set by pressing and holding
g
+
t
.
Press and hold
g
+
t
This is not indicated on the front panel, so it can be
used to prevent unlocking when not desired.
Press and hold
g
+
t
to unlock this key lock.
This can also be used to unlock when key lock has
been set by pressing and holding ESC+ENT.