1-3. Sensor signal input terminals
1
2
3
4
5
6
7
+EXC(A)
–SIG(B)
–EXC(C)
+SIG(D)
SHIELD
+SENS/TEDS(F)
–SENS/GND(G)
Terminal
number
Signal name
Wire color
1
+SENS/TEDS (F)
Orange
2
−SENS/GND (G)
Green
3
+EXC (A)
Red
4
−SIG (B)
Black
5
−EXC (C)
Blue
6
+SIG (D)
White
7
SHIELD (E)
Yellow
o
Terminals 1 and 2 are data terminals for use by both remote
sense and TEDS sensors. Before connecting the sensor,
set which type will be connected in advance using the
Remote Sense/TEDS setting on the Calibration menu
page. If “Remote Sense disabled/TEDS enabled” is selected,
calibration will occur automatically if a sensor with TEDS
built-in is connected.
o
The wire colors are those that we use in the strain gauge
transducers that we make.
1-4. D/A output connectors
These can output either voltage or current.
They cannot output both at the same time. Set voltage or
current in System Settings
q
D/A Converter
q
D/A Output
Mode (page 75).
Terminal
number
Signal
name
Explanation
9
V-OUT
D/A voltage output
10
I-OUT
D/A current output
11
COM
D/A output common terminal
o
The D/A output is isolated from this unit’s circuits.