TE Technology TC-24-25 Скачать руководство пользователя страница 29

 
 

29

 

31.

 

EEPROM WRITE ENABLE 

  Write 

Command: 

34 

hex 

  Read 

Command: 

 

4c 

hex 

  Interpret: 

 0 

is 

disable 

EEPROM 

writes 

     1 

is 

enable 

EEPROM 

writes 

Note: 

This function is always stored to EEPROM.  This feature is 
available for controllers whose EEPROM is revision D or later.  
(The EEPROM has a white sticker on it labeled “C74-114D” where 
the last letter indicates the revision level.) 

Function Description: Upon a power-up or reset condition, the controller performs an 

initialization of all command variables that have write commands by 
transferring the last values stored in non-volatile memory 
(EEPROM) to appropriately referenced static RAM locations.  This 
action is performed so that the controller can run faster (RAM is 
faster than ROM).  When the “EEPROM WRITE ENABLE” is 
enabled, any changes in the run-time values of the command 
variables are also stored in EEPROM as well in RAM and thus will 
be recalled upon a power-up or reset condition.  When 
the “EEPROM WRITE ENABLE” is disabled, run-time values are 
stored only in RAM.  This setting gives you the ability to change 
run-time values without changing your desired power-up settings.  
This will also alleviate a possible problem since the maximum 
number of writes to an EEPROM location is 1,000,000. 

 

V.  ASCII Reference Table 

 
 Dec   Oct   Hex   Binary          Description 
  
 000   000   00   00000000   ^@ ^` NULL NUL null c-@ c-` 
 001   001   01   00000001   ^A ^a SOH GTL c-A c-a start-of-heading 
 002   002   02   00000010   ^B ^b STX c-B c-b start-of-text 
 003   003   03   00000011   ^C ^c ETX c-C c-c end-of-text 
 004   004   04   00000100   ^D ^d EOT SDC end-of-transmission c-D c-d ..._._ 
 005   005   05   00000101   ^E ^e ENQ PPC c-E c-e enquiry 
 006   006   06   00000110   ^F ^f ACK c-F c-f acknowledge 
 007   007   07   00000111   ^G ^g BELL BEL bell c-G c-g \a 
 008   010   08   00001000   ^H ^h BS GET backspace c-H c-h \b 
 009   011   09   00001001   ^I ^i TAB TCT HT tab c-I c-i \t 
 010   012   0A   00001010   ^J ^j LF lf linefeed c-J c-j \n 
 011   013   0B   00001011   ^K ^k VT vertical-tab c-K c-k \v 
 012   014   0C   00001100   ^L ^l FF ff formfeed page \f c-L c-l 
 013   015   0D   00001101   ^M ^m CR cr carriage-return c-M c-m \r 
 014   016   0E   00001110   ^N ^n SO c-N c-n shift-out 
 015   017   0F   00001111   ^O ^o SI c-O c-o shift-in 
  
 016   020   10   00010000   ^P ^p DLE c-P c-p data-link-escape 
 017   021   11   00010001   ^Q ^q DC1 LLO go XON xon c-Q c-Q 
 018   022   12   00010010   ^R ^r DC2 c-R c-r 
 019   023   13   00010011   ^S ^s DC3 stop XOFF xoff c-S c-s 
 020   024   14   00010100   ^T ^t DC4 DCL c-T c-t 
 021   025   15   00010101   ^U ^u NAK PPU negative-acknowledge c-U c-u 
 022   026   16   00010110   ^V ^v SYN c-V c-v synchronous-idle 

Содержание TC-24-25

Страница 1: ...4 25 RS232 Thermoelectric Cooler Temperature Controller November 8 2005 Drawing 3599 Rev F TETECHNOLOGY INC 1590 Keane Drive Phone 231 929 3966 Traverse City MI 49686 8257 Fax 231 929 4163 http www te...

Страница 2: ...ng of the thermoelectric device either of which are known to reduce the lifetime of any thermoelectric device Care should be taken to prevent the temperature of the thermoelectric device from going be...

Страница 3: ...able Computer set with controller stand alone operation Remote user set temperature potentiometer 0 mA to 20 mA current loop 0 VDC to 5 VDC adjustable range Differential temperature control No compute...

Страница 4: ...ol Mechanically the control printed circuit board is mounted to a metal bracket that is suitable for either horizontal or vertical orientation Input and output connections are accessible via screw typ...

Страница 5: ...ature All temperatures may be consistently displayed in C or F Two types of control output modes may be selected This determines the direction of the current flow through the thermoelectric cooler dur...

Страница 6: ...r supplied with the TC 24 25 RS232 is the MP 2379 Appendix D Thermistor Styles for TC 24 25 RS232 shows dimensional data on the MP 2379 and other thermistor styles readily available as well as tempera...

Страница 7: ...en click the Send Box Values button prior to proceeding 2 2 First select the PWM OUTPUT TIMEBASE menu key Options for SLOW TIMEBASE 675Hz or FAST TIMEBASE 2700Hz will appear Select the option that is...

Страница 8: ...tions NO ALARM PICKED indicates that no alarm parameters are desired SET TRACKING ALARMS allows an alarm to be set with respect to the set temperature It will move accordingly with a change of the tem...

Страница 9: ...C OR DEG F menu key permits the selection of displaying F or C 2 14 The EEPROM WRITE ENABLE located just below the CONFIGURE SECTION can be checked or unchecked depending on your requirements See Appe...

Страница 10: ...Send Box Values button to download these constants to the controller 4 0 Calibration 4 1 The CALIBRATE section provides additional variables that can be used to fine tune your system s operation You...

Страница 11: ...stom Software The controller does not specifically have to be used with the software that is provided You could also use commercial software such as LabVIEW from National Instruments This might be use...

Страница 12: ...Controller Wiring Diagram 12...

Страница 13: ...uous current 8 7 6 5 4 3 2 1 Expansion Connector Molex 22 23 2081 25 mA Current Source Alarm Cancel Input Common High Alarm Low Alarm No Alarm JP2 8 7 6 5 4 3 2 1 Customer supplied Connector Note Pins...

Страница 14: ...Mechanical Package Drawing 14...

Страница 15: ...O 100 C 10K OHM POTENTIOMETER 0 TO 5 VDC 4 to 20 mA DC OR 0 TO 20 mA DC PC SET VALUE DIFF SET INPUT 2 INPUT 2 AUXILLARY INPUT CONTROLLER INPUT IMPEDANCE 6 5K OHMS REQUIRES ADDITION OF EXTERNAL RESISTO...

Страница 16: ...WIRE TO JP3 3 BLACK WIRE TO JP3 3 RED WIRE TO JP3 2 RED WIRE TO JP3 2 CLEAR WIRE TO JP3 1 CLEAR WIRE TO JP3 1 THERMISTOR WIRES TO JP1 5 AND JP1 6 POLARITY NOT IMPORTANT THERMISTOR WIRES TO JP1 5 AND J...

Страница 17: ...owing table offers possible solutions corresponding to a particular error Error Message Cause Solution Comm Port Timeout No power to TC 24 25 RS232 unit Apply power to TC 24 25 RS232 review customer d...

Страница 18: ...hand if the bandwidth is too small and there is any time lag between the cooling heating element and the temperature sensor the temperature will initially overshoot the set point and then settle into...

Страница 19: ...compensate for rapid changes in system loading This term is generally used only on very sluggish systems or where very quick response is necessary The acceptable range for the TC 24 25 is 0 1 to 10 cy...

Страница 20: ...presented by two hex characters is compared to the two ASCII character hex checksum sent by the controlling computer The TC 24 25 RS232 will respond as shown below depending on whether or not the chec...

Страница 21: ...lue of 0 selects a computer communicated set value so the eight character send value DDDDDDDD is 00000000 4 Compute the checksum SS by adding the ASCII values of the following characters 0 0 2 9 0 0 0...

Страница 22: ...will send back stx 000003e8c0 ack If the checksum is not correct the temperature controller will send back stx XXXXXXXXc0 ack Read the actual temperature of the control thermistor INPUT1 from the con...

Страница 23: ...on One millisecond might be a good delay time to use initially but the exact time will vary depending on the host computer s hardware and the particular demands on the controller at the moment IV Comm...

Страница 24: ...This function tells the controller how the set point temperature will be communicated 0 sent or returned means computer communicated set value 1 sent or returned means Potentiometer Input 2 sent or re...

Страница 25: ...25 Read Command 45 hex Interpret 0 is HEAT WP1 and WP2 1 is HEAT WP2 and WP1...

Страница 26: ...lue to decimal and divide by 10 0 to convert to F C 13 PROPORTIONAL BANDWIDTH Write Command 1d hex Read Command 51 hex Interpret Fixed point temperature bandwidth in F C Multiply desired bandwidth by...

Страница 27: ...Temperature reference to compare against Input1 for high alarm output 20 LOW ALARM SETTING Write Command 24 hex Read Command 58 hex Interpret Temperature reference to compare against Input1 for low al...

Страница 28: ...command to reset the alarm latches 27 HEAT MULTIPLIER Write Command 0c hex Read Command 5c hex Interpret This multiplies the heater percentage of power to offset its effectiveness 100 is a multiplier...

Страница 29: ...will also alleviate a possible problem since the maximum number of writes to an EEPROM location is 1 000 000 V ASCII Reference Table Dec Oct Hex Binary Description 000 000 00 00000000 NULL NUL null c...

Страница 30: ...0101010 asterisk star 043 053 2B 00101011 addition sign 044 054 2C 00101100 comma 045 055 2D 00101101 subtraction sign minus hyphen negative dash 046 056 2E 00101110 period dot decimal 047 057 2F 0010...

Страница 31: ...01100001 a 098 142 62 01100010 b 099 143 63 01100011 c 100 144 64 01100100 d 101 145 65 01100101 e 102 146 66 01100110 f 103 147 67 01100111 g 104 150 68 01101000 h 105 151 69 01101001 i 106 152 6A 01...

Страница 32: ...APPENDIX D Thermistor Styles for TC 24 25 RS232 MP 2444 MP 2444 MP 2379 MP 2379 MP 2379 MP 2542 MP 2542 MP 2542 32...

Страница 33: ...40 7976 61 3590 82 1761 1 51750 20 18751 41 7662 62 3464 83 1705 0 49157 21 17923 42 7362 63 3343 84 1652 20 146735 1 46709 22 17136 43 7075 64 3227 85 1601 19 138447 2 44397 23 16388 44 6801 65 3115...

Отзывы: