background image

NOTE

i

408-1542

3

 of 6

Tyco Electronics Corporation

Rev 

N

Figure 3

5.2. Lubrication

Lubricate all pins, pivot points, and bearing surfaces 

with SAE 20 motor oil as follows:

Tools used in daily production - lubricate daily

Tools used daily (occasional) - lubricate weekly

Tools used weekly - lubricate monthly

Wipe excess oil from tool, particularly from crimping 

area. Oil transferred from the crimping area onto 

certain terminations may affect the electrical 

characteristics of an application. 

5.3. Periodic Inspection

1. The tool should be immersed (handles 

partially closed) in a reliable commercial 

degreasing compound to remove accumulated 

dirt, grease, and foreign matter.

2. Close tool handles until ratchet releases and 

then allow them to open freely. If they do not 

open quickly and fully, the spring is defective and 

must be replaced. See Section 6, 

REPLACEMENT AND REPAIR.

3. Inspect head assembly for worn, cracked, or 

broken jaws. Refer to Figure 5. If damage is 

evident, return the tool to TE for evaluation and 

repair. See Section 6, REPLACEMENT AND 

REPAIR.

5.4. Gaging the Crimping Chambers

This inspection requires the use of a plug gage 

conforming to the dimensions provided in Figure 6 

and Figure 

TE does not manufacture or market these gages. To 

gage the crimping chambers, proceed as follows:

Remove traces of oil or dirt from the crimping 

chambers and plug gage.

4. Close the tool handles until it is evident that 

the jaws have bottomed; then hold in this 

position. Do NOT force the jaws beyond initial 

contact.

5. Align the GO element with the crimping 

chamber. Push element straight into the 

crimping chamber without using force. The GO 

element must pass completely through the 

crimping chamber.

6. Align the NO-GO element and try to insert it 

straight into the same crimping chamber. The 

NO-GO element may start entry, but must not 

pass completely through the crimping chamber.

The jaws in Tools 49935 and 574573 are not 

symmetrically located; make sure that the tool 

conforms to the symmetry requirements shown in 

Figure 8.

If the crimping chambers conforms to the gage 

inspection, the tool is considered dimensionally 

correct, and should be lubricated with a THIN coat of 

any good SAE 20 motor oil. If not, the tool must be 

returned for further evaluation and repair. Refer to 

Section 6, REPLACEMENT AND REPAIR.

For additional information regarding the use of a plug 

gage, refer to Instruction Sheet 408-7424.

Terminals and Splices Can Also Be Rotated 180° with the Seam Toward the Nest

Отзывы: