-3-
TDK-Lambda
HWS1000/ME Series
INSTRUCTION MANUAL
Special Instructions for IEC/EN/UL/CSA 60601-1
·
These products are designed for continuous operation within an overall enclosure, and must be mounted such that access to the
mains terminals is restricted. See Clause 16, IEC/EN/UL60601-1.
·
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with oxygen or with nitrous
oxide.
·
These products are classed as ordinary equipment according to IEC/EN/UL60601-1 and are NOT protected against the ingress of
water.
·
Connect only apparatus complying with IEC/EN/UL60601-1 to the signal ports.
·
Except for permanently installed equipment as defined in Clause 57.6 of IEC/EN/UL60601-1 the overall equipment in which these
products are installed must have double pole fusing on the input mains supply.
The products themselves have single pole fusing in the live line.
·
These products provide the following insulation between mains and output.
(a) As for UL60601-1 : 2006 (1st Ed.), EN60601-1 : 1990+A1+A2+A13 and CSA C22.2 No.601-1:M90, 2005,
basic insulation provided. Sure to add supplemental insulation to input or output in the equipment.
(b) As for ANSI AAMI/ES60601-1 : 2005, EN60601-1 : 2006 and CSA C22.2 No.60601-1 : 08, reinforced
insulation (2MOOP) provided.
·
When the PSU is installed within medical equipment an all pole mains input disconnect device must be fitted.
·
Reference should be made to local regulations concerning the disposal of these products at the end of their useful life.
·
The maximum normal leakage current of this product is 500 microamperes for IEC/EN/UL60601-1.
When using it as a patient care
equipment, all outer surfaces of the equipment shall be constructed of nonconductive material.
See Clause 19.5DV.2 of UL60601-1.
·
These products have not been assessed to IEC/EN60601-1-2 (EMC) but EMC test data is available from TDK-Lambda Corp..
Note
·
Take note that traces of sheet metal processing be left in our power supplies.
·
When disposing product, follow disposal laws of each municipality.
·
Published EMI (CE, RE) or immunity is the result when measured in our standard measurement conditions and might not satisfy
specification when mounted and wired inside end-user equipment.
Use the product after sufficiently evaluating at actual end-user equipment.
·
When exporting our products, apply for necessary permissions as required by rules and regulations of Foreign Exchange and
Foreign Trade Control Act.
·
Catalogue, contents of the instruction manual may be changed without a prior notice. Refer to latest catalogue or instruction manual
·
Reproduction or reprinting the instruction manual or its portion is forbidden without our permission.
LONG-TERM STORAGE METHOD AND LONG-TERM STORAGE PERIOD
·
Please keep the product in carton box.
·
Please do not apply excessive vibration, shock or mechanical stress applied directly to the product.
·
Please keep away from direct sunlight.
·
For long-term storage temperature and humidity, the following conditions shall be used as a guideline :
Temperature range : 5℃
~
30℃
Humidity range : 40%
~
60%RH
Please keep away from the places where temperature and humidity can change drastically.
It can cause condensation on the product or deterioration.
·
For long-term storage period, we recommend to use within 2 years after receiving the product.
There is tendency that the leakage current of an aluminum electrolytic capacitor may increase when stored without using for
a long time. This phenomenon can be improved by applying voltage to the aluminum electrolytic capacitor to reduce the
increased leakage current through the self-recovery effect of the electrolyte. For reference, before using products that have been
stored for a very long time, please warm-up first for 30 minutes or more without taking load.
< Criterion of warm up voltage condition >
(1)Implementation period : 1 year or above after the delivery
(2)Electrical continuity condition
Input voltage : Rating
Load : 0A
Ambient temperature : Normal temperature
Time : 30 minutes or more