TDK-Lambda 

CUS800M 
INSTRUCTION MANUAL

 

 

Special Instructions for IEC/EN/ES/CSA 60601-1 

 

These products are designed for continuous operation within an overall enclosure, and must be mounted such that access to the mains 

terminals is restricted. See Clause 16, IEC/60601-1 2

nd

 edition or clause 8 IEC/EN/ES/CSA 60601-1 3

rd

 edition. 

 

These products are NOT suitable for use in the presence of flammable anesthetics mixtures with air or with oxygen or with nitrous 

oxide. 

 

These products are classed as ordinary equipment according to IEC/EN/ES/CSA60601-1 and are NOT protected against the ingress 

of water. 

 

Reference should be made to local regulations concerning the disposal of these products at out of their useful life. 

 

These products have not been assessed to IEC/EN60601-1-2 (EMC) but EMC test data is available from TDK-Lambda Corporation.

 

 

For IEC 60601-1 2

nd

  Edition, these products have a reinforced  insulation barrier between input and output. For IEC/EN/ES/CSA 

60601-1 3

rd

 edition, these products provide reinforced insulation between input and outputs of 2 MOPP, 1 MOPP from input to earth 

and 1 MOPP from output to earth. 

 

These products have SELV outputs. 

 

All outputs have basic spacing’s to earth rated for mains – 250Vac, and due consideration must be given to this in the end product 

design. 

Important safety instructions 

 

Servicing 

 

These products are not customer serviceable. Repairs may only be carried out by TDK-Lambda  Corporation  or their authorized 

agents. These products are not authorized for use as critical components in nuclear  control systems, life support systems or 

equipment for use in hazardous environments without the express written approval of the Managing Director of TDK-Lambda 

Corporation. 

 

Safety Class of Protection 

These products are designed for the following parameters: Material Group IIIb, Pollution Degree 2, Over voltage Category II, class I 

(earthed)   

Indoor use as part of an overall equipment such that the product is accessible to service engineers only. 

   

EMC performance 

 

Immunity (IEC61000-6-2) 

 

Test 

Standard 

Electrostatic discharge 

IEC61000-4-2 

Electromagnetic field 

IEC61000-4-3 

Fast / burst transient 

IEC61000-4-4 

Surge immunity 

IEC61000-4-5 

Conducted RF immunity 

IEC61000-4-6 

Power frequency magnetic field 

IEC61000-4-8 

Voltage dips, variations, interruptions 

IEC61000-4-11 

Radiated Field In Close Proximity 

IEC61000-4-39 

 

 

Emissions 

 

            Test 

Standard 

Comments 

Radiated electric field 

EN55032 

Class B (as per CISPR 32) 

Conducted emissions 

EN55032 

Class B (as per CISPR 32) 

Conducted harmonics 

IEC61000-3-2 

Compliant 

 

 

 

 

 

Содержание CUS800M

Страница 1: ...e to high voltage and high temperature You might also get electric shock or burn Do not make unauthorized changes to this product nor remove the cover as you might get an electric shock or might damag...

Страница 2: ...re there is conductive foreign matter or dust there is possibility of product failure or malfunction Provide countermeasure for prevention of lightning surge voltage as there is risk of damage due to...

Страница 3: ...uts have basic spacing s to earth rated for mains 250Vac and due consideration must be given to this in the end product design Important safety instructions Servicing These products are not customer s...

Страница 4: ...ange 5 C 30 C Humidity range 40 60 RH Please keep away from the places where temperature and humidity can change drastically It can cause condensation on the product or deterioration For long term sto...

Страница 5: ...lank Standard type SF Single fuse model 2 Terminal Explanation Input side Top side L Input terminal Live line Fuse in line M3 screw N Input terminal Neutral line Fuse in line M3 screw Functional earth...

Страница 6: ...e ON OFF terminal Please refer to 6 14 8 R Remote ON OFF terminal Please refer to 6 14 TB1 CN51 CN61 Connector Housing Terminal Pin Connectors used Matching housings Matching pins Maker Input TB1 T825...

Страница 7: ...ain Output ON OFF Control V DC_OK detection S S R R Fuse rating 16A Circuit topology switch frequency PFHC circuit active filter 65kHz Typ Output circuit LLC resonant 50k 300kHz Standby supply circuit...

Страница 8: ...me chart 2 Input voltage Standby Supply Remote ON OFF 2 Main Output PG signal OCP act on 0V 0V R 0V 0V Remote ON Remote OFF 100ms OCP reset DC_OK monitor point R Remote ON OFF 2 Standby Supply off OCP...

Страница 9: ...nts when connecting or removing connectors Remote ON OFF control lines shall be twisted or used shielded wires Separate from load line Remote sensing lines shall be twisted or used shielded wires End...

Страница 10: ...OFF and then remote ON again Refer to 6 14 6 5 Over Temperature Protection OTP Over temperature shut down function is provided When ambient or internal temperature rises abnormally over temperature p...

Страница 11: ...ossible Power Supply Power Supply 4 Please select a reverse current prevention diode with maximum forward current rating more than output load current V V V V 4 6 10 Isolation Test Isolation resistanc...

Страница 12: ...sofoutputvoltage since ACfailthrough PGand STBY Power Good signal shows High level 4 5V 5V to indicatepowersupply operating normal and can sourcemaximum2mAcurrent When powersupply stopsoperating thesi...

Страница 13: ...inals to load terminals Connect S terminal to terminal of load and S terminal to terminal of load with sensing wires The total line voltage drop side line and side line shall be less than 0 5V In case...

Страница 14: ...should be considered Please refer to the output derating versus input voltage curve 6 1 for detail Main Output Derating versus Operating Ambient Temperature Ta Main Output Power Derating versus High O...

Страница 15: ...to 4 secondsis allowed T1 4 s 3 Maximumduty ofpeak powershould belessthan 40 1 1 2 40 7 3 Mounting Method 1 These models are forced air cooling type power supply The power supply has ventilating hole...

Страница 16: ...dered by in rush current value at line turn on Do not select the fuse according to input current RMS values under the actual load condition CUS800M series F16AH 11 Before concluding that the unit is a...

Страница 17: ...ance with the Low Voltage Directive EMC Directive and RoHS Directive UKCA MARKING UKCA Marking when applied to a product or packing material for a product covered by this handbook indicates compliance...

Отзывы: