TDI International, Inc.
The Technical Source!
Page 26
TECHNICAL SPECIFICATIONS
AC Test Mode Test Frequency:
1kHz, 10kHz, 120Hz, 100Hz
Test Frequency Accuracy:
50 PPM (0.005%)
Test Signal Level:
0.25 / 0.5 / 1.0 +/- Vrms Sine Wave
Source Impedance:
62.5 / 1k / 16k +/- 1%
TYPICAL OFFSET
Resistance
≤ 25m
Capacitance
0.65 pF
Inductance
0.1 uH
Offset value should be subtracted from measurement result for small value compo-
nents (R<10
, C<100pG, L<10
H).
MEASUREMENT RANGES
Parameter
Measurement Range
Test Frequency
Resistance
<9.9 M
1kHz
Capacitance
<9999pF
10kHz
10000pF to 1
F
1kHz
>1uF
100Hz
Inductance
0.5
H to 99 H
10kHz
100
H 99mH
1kHz
>100mH
100Hz
APPENDIX A. SPECIFICATIONS