INSTRUCTION
MANUAL
- MODEL
DLP 400-5-750A
2
2.0
SPECIFICATIONS
The following
ratings
apply:
LOAD
VOLTAGE:
10
TO
400V
LOAD
CURRENT: 0
TO
5A
POWER
DISSIPATION:
0
TO 750W
OVERLOAD
RATING:
10
%
SELF-PROTECTION
—
OVERVOLTAGE:
<500V
OVERCURRENT:
<7
A
OVERPOWER:
<1000W
2 .1
MODE
SELECTOR SWITCH
POSITION
(From
left
to
right)
POSITION
1
:
POSITION
2
:
POSITION
3
:
POSITION 4
:
POSITION
5
:
POSITION
6:
POSITION
7
:
POSITION
8
:
Constant
resistance
0-.01 A/V
as
determined
by the
front
panel
DC
load
adjust
Constant
resistance
0-.03
A/V
as
determined
by
the
front
panel
DC
load
adjust
Constant
current
0-1A
as
determined
by
the
front
panel
DC
load
adjust
Constant current 0-5A
as
determined by
the
front
panel
load
adjust
Constant
voltage
load. In
this
position,
the
load
is
similar to
a
battery
being
charged
or
a
constant
voltage
zener
diode;
no
current
is drawn
until
the
source voltage
reaches
the regulating
voltage.
The
voltage
at
which
the
Dynaload
regulates
is
adjustable by the
front
panel
volts
control.
External
modulation
—
will
program
from
0-5A
with
0-6V
applied to
the
external
modulation
terminals
(TB1)
.
Modulation
sensitivity
is
directly
adjustable
by
the
front
panel
DC
load
adjust
control.
Pulse
—
A
square
wave
pulse
lo<iji which
may
be
varied
from
0-5A
and 500-5000
Hz
as
desired
by
the
front
panel
pulse
controls
.
Short —
A
short
circuit
which
is
applied
in
series
with
the ammeter
across
the
input
allowing
short
circuit
current
tests
.
[ELP400-5-750A]