3. Description

The measurement modes volt AC/DC, ampere AC/DC, frequency and 5 A current
transformer are supported. Scale endpoint of the indicator as well as scaling of the
analogue signal and current transformer measurement can be adjusted freely via
the DIP switch, alternatively there is an option for automatic adaption of the scale
endpoint. The unit can be switched between volt and ampere measurement via an
external switch. Minimum and maximum values are recorded and can be displayed
optionally using an external switch. Measured values are recorded over a time
span of 3 minutes up to 14 days. The time base as well as the display of the
measuring history can be activated by an external switch. The measured values
remain stored as long as the device is supplied with voltage.
The following parameters can be adjusted using the USB software: scale endpoint,
scale caption,  display  style  (pointer,  tachometer,  bar graph,  and  more),  splash
image, thresholds for alarm output, hysteresis, and more. Recorded measurement
values can be read and a continuous transmission of the actual measuring value
can be enabled. Using the software, values can be viewed and exported as CSV.

Front

1

Unit or free text

2

Threshold values

3

Upper scale caption

4

Graphical measurement display

5

Digital measurement value

6

Lower scale caption

Back

1

2 optocoupler switch outputs

2

DIP switch for setting measurement 
mode and upper scale caption

3

Supply voltage input

4

Jumpers for configuration or 
connection external switch for 
activating/switching measurement 
mode, graphical historic data display,
min/max display, auto-scaling

5

USB interface for alternative power 
supply and software communication 
interface

6

Measurement inputs

----- 

2

 -----

Содержание 101398

Страница 1: ...working on the device hazardous voltages must not be connected to the device The device must not be used as the only protective device or protective shutdown Read instruction manual carefully before...

Страница 2: ...ice is supplied with voltage The following parameters can be adjusted using the USB software scale endpoint scale caption display style pointer tachometer bar graph and more splash image thresholds fo...

Страница 3: ...nt transformer voltage direct Load L N Current Transformer max 500 V Ilow Ihigh Uhigh Ulow Current on high side 50 V max Load L N max 50 V Ilow Ihigh Uhigh Ulow Current on high side voltage direct 50...

Страница 4: ...er Configuration via USB interface with software Digalox Manager For configuration with software set DIP switch 10 to OFF position Please note that the measurement mode selected in Digalox Manager is...

Страница 5: ...repeat steps 2 8 until all required measurement modes are configured The time base does not need to be set again as it is used for all measurement modes 10 In operation the measurement modes can be c...

Страница 6: ...lt AC ON 1 2 3 4 5 6 7 8 9 10 11 12 J1 J6 ON 1 2 3 4 5 6 7 8 9 10 11 12 J1 J6 A AC direct Ampere AC ON 1 2 3 4 5 6 7 8 9 10 11 12 J1 J6 ON 1 2 3 4 5 6 7 8 9 10 11 12 J1 J6 V DC direct Volt DC ON 1 2 3...

Страница 7: ...ing and closing J1 the current time base is displayed For each subsequent opening and closing the time base changes to the next setting In order to save the setting permanently DIP switch 10 must be s...

Страница 8: ...e 2 6 M Measuring range ampere 10 A AC DC and 5 A AC for current transformer Accuracy ampere 1 true RMS Internal resistance voltage 5 m Measuring range frequency 10 500 Hz Accuracy frequency 0 1 Hz Re...

Отзывы: