Shellsol D70-Exxol D80-Sepisol KD70 

190-257 

Shellsol A Solvesso 100 Sepisol A 

162-180 

Shellsol AB-Solvesso 150 

182-215 

Shellsol H-Varsol 60-Sepisol H 

180-210 

Toluene Toulol 

     111 

     535 

Turpentine Gum Spirits 

            152-270 

    250 

Xylol/Xylene                                                                                      140                 525  

 

 

IMPORTANT INFORMATION:   Never open the recycler prior to the suggested cool of time of 50° C.   

Always be certain that the cover gasket does not show signs of wear or have any debris attached to it.   The 

green "on" light should be indicated prior to starting each cycle. 

 

WARNING:   Never change the thermal oil before allowing the machine to cool for at least 48 hours. 

 
Nonflammable Solvents 

 Boiling 

Working 

Name Point 

(°C) Temp.(°C) 

 

Chloroform-Trichloromethane 661 

90 

Methilene Chloride-DiChlormethane 

40 

70 

Freon 113-Frigen 113 

46 

80 

Carbon Tetrachloride-Tetrachloromethane 

78 

110 

1.1.2.2 Tetrachloroethane Acetylene Tetrachloride 

147 

160 

Tetrachloroethylene-Pestabil-Vors Perchloroethylene-Per 

121 

150 

Trichloroethane-1.1.1.3*1- Chlorothene-Genklene-Solvetane 

Baltane-Mecloran 74 

110 

Tricholoethylene-Tri Vorclin-Althene-Triklone 

87 

120 

 
 
Warnings: 

 

1. When treating nonflammable solvents do not set the working thermostat at a temperature higher than that 

shown above.   Higher temperatures may produce hydrochloric acid, acidify the solvent and consequently 

damage the recycler.  

 

2. Any solvent containing more than 35% water will not recycle successfully. 
 

 
 
MAINTENANCE 
 
Daily -          

A)  Clean the distillation vessel …..remainder of 8.1 from Omega Manual 

                      B)  it is best to remove paraffin build-up when the unit is warm 

 

 

Weekly

 -   A)  Using a soft cloth and mild solvent clean up the exterior of the   

                           apparatus. 

 

C) 

 

Check the thermal oil level when the unit is cold, the level should be at    

      the lower indicator.    

 

Содержание SR12

Страница 1: ...Triangle Biomedical Sciences Inc Solvent Recovery System Operation And Service Manual SR12 Version 1 2 Serial ___________________...

Страница 2: ...ION 4 Application 4 Function 4 Separation Module 4 Safety Measurers 4 OPERATIONAL SAFETY 4 Location 4 Solvent Containment 4 Safe Exposure Temperature 5 OPERATING PROCEDURES 6 Purpose 6 Procedure for A...

Страница 3: ...Page 3 GLOSSARY 15...

Страница 4: ...ansfer the benefit of such warranties to the user This warranty supersedes and is given in lieu of all implied warranties and is void if user does not provide unit with continuous ample electrical pow...

Страница 5: ...illation using the Separation Module and Baffle The Switching Module allows an operator to set a fraction temperature at which the distillate of the different solvents will be separated The fraction t...

Страница 6: ...on the floor under your recovery unit This pan is recommended for use as a safety pan to prevent solvents from spilling onto the floors in case of an over flow due to operator error Appropriate 3 gal...

Страница 7: ...taminated solvent in the distilling chamber up to 3 gallons NOTE Take special care NOT to allow any contaminated solvent to enter the Vapor Collector Duct Manifold when you are filling the distilling...

Страница 8: ...hen operating This could create a dangerous situation Xylene 1 Place clean and dry collection containers under each of the lower distillate solvent outlet nozzles left and right 2 Verify that the Ther...

Страница 9: ...overy cycle has been completed let the unit cool to at least 50 C prior to opening the secured top Premature opening will allow a release of solvent fumes into the surrounding atmosphere and can also...

Страница 10: ...156 377 Cyclohexane 81 260 Cyclohexanon 155 419 1 2 DiChlorethane 84 412 DiEthyl Cellosolve 121 208 DiEthyl Ketone 104 445 Ethanol 82 400 Ethyl Acetate 79 427 Ethyl Amyl Keytone EAK 160 Ethylene Glyc...

Страница 11: ...113 46 80 Carbon Tetrachloride Tetrachloromethane 78 110 1 1 2 2 Tetrachloroethane Acetylene Tetrachloride 147 160 Tetrachloroethylene Pestabil Vors Perchloroethylene Per 121 150 Trichloroethane 1 1...

Страница 12: ...o control the temperature of the thermal oil that transfers heat to the distilling chamber Range 40 200 C Note The oil temperature is generally 15 20 C higher than the temperature of the contaminated...

Страница 13: ...time lasts much longer than 4 hours 1 Contaminated solvent contains too much water 2 Thermal oil level is low 3 Thermal oil is old and dirty 4 Heating elements need cleaning 1 Replace the solvent 2 Ad...

Страница 14: ...er 11 Cooling Fan Motor Expl Proof 12 3 Way Fraction Switch Expl Prf 13 Heater Cable Sheath Pipe Expl Prf 14 Solvent Receiver Tanks 15 Solvent 16 Oil Discharge 17 Thermal Oil Level Indicator 18 Workin...

Страница 15: ...g Thermostat T1 384010 Safety Thermostat T2 465000 White Warning Light 465001 Red Warning Light 465002 Green Warning Light 475000 Ventilator Motor 468402 Heating Element 385007 Timer 425002 Control Kn...

Страница 16: ...since the distlland mixture is forced into the area in which only pure material is desired Fraction A portion of the collected separated material recovered from the head of the still Fractional Disti...

Отзывы: