- 8 -
Factory settings
Language on screen:
English
or
German
Application time per frequency:
3:00 Minutes
(adjustable betw een 0:30 and 59:59)
Signal strength:
80%
of maximum of 12Vpp
(adju/- in steps of 1%)
Signal form:
AUTO
(Sine, Square Full Wave, Half Wave, SMART and OFF
selectable). AUTO = Square Half Wave if low er than 65000 Hz, Sine
above 65000 Hz.
Memory of ”123“ is empty.
Memory of “DUO+” is e mpty
Memory of ”ABC“ is empty and locked.
Connector 3.5mm CON1 :
At this connector is the standard multi signal output available w ith a bandw idth up to 3 MHz. Standard
application parts may be connected.
Connector 3.5mm CON2 :
At this connector is the WIDE output available w ith a bandw idth up to 15 MHz. Standard application parts may
be connected.
Connector 2.5mm on CON3 :
Connector f or the Pow erPort. Application parts such as a laser pointer or magnetic field electrode must b e
connected here. No standard application part may be connected.
RESET: If device cannot be operated from the touch screen due to unknow n reasons, the RESET button on
the rear panel can be pressed shortly to shut dow n the unit. Use a “tooth pick” or “match” carefully.
Such effects may happen if the battery has not been charged correctly or a data transfer from sof tw are was
interrupted.
On top of device: 5 plugs red, black, green, yellow and blue (return/ground) for the 4 channel applications.