Instruction Manual
®
Leading the Way in Accuracy
1442
CP1442-7.08R1
www.taylorusa.com
Thank you for purchasing a Taylor® Recording Thermometer
with dual probes. This product is an example of superior
design and craftsmanship. This product will the allow user to
record and monitor temperature and would be ideal for most
hospital environments. Please read this instruction manual
carefully before use. Keep these instructions handy for future
reference.
Description of Parts
Description of Parts
Battery Installation
1.
LCD
read out
2.
Right sensor port
3.
Left sensor port
4.
Left '
SET
' button -
For setting target temperature set points
;
confirming setting
operation
;
scrolling
alarm
records.
5.
Left '
UP
' button -
For scrolling the alarm events
;
increasing
the
displayed
value.
6.
Left '
DOWN
' button -
For scrolling the alarm events
;
decreasing
the
displayed
value
.
7.
'
Pass Word
' button -
For accessing the unit, new password
setting
and turning
off
the
alarm
sounds
.
8.
Velcro holding and mounting pads
9.
Battery compartment cover
10.
'
Password
' reset button
11.
Right, left and dual display switch
12.
'
System Reset
' button
13.
Battery placement recess
14.
Sensor Probes
The
operates on (2) AAA batteries.
1.
Use a small phillips head screw driver to loosen the 6
small screws that hold the battery compartment cover in
place.
2.
Remove the compartment cover.
3.
Install (2) AAA batteries into battery recess. Place the
negative side of the batteries with the right side orientation
for both as show in the
Figure #1
.
4.
Replace the compartment cover and tighten screws.
Taylor® Recording Thermometer with dual probes
1.
There are two individual removable probes and jacks on
either side of the instrument. Remove protective cap from
jacks and plug in probes. Make sure Right-Left-Dual display
switch is in the correct position before turning on
thermometer.
2.
There is one display with a two line readout on the unit.
The top line is for temperature display. The left area is for
left probe, the right area for right probe. The bottom line
is for displaying the clock (24 hour format only), and the
status of the probes or the alarm event records. The right
part of the bottom line “88H88M88S” is also for alarm
delay time setting. For example, when the delay time is set
to “00H15M30S” and the temperature exceeds the limit
that the user set, the alarm will not sound unless the
temperature has exceeded the set temperature for 15
minutes 30 seconds. The delay time can be set from
00H00M00S to 10H59M59S (from 0 Hour 0 Minute 0
Second to 10 Hour and 59 Minute 59 Second). If the
temperature exceeds the set temperature but does not stay
there for longer than the set delay, then the sound alarm
will never occur but the information is still recorded. The “L”
is for showing that the data displayed is for the left probe;
the “R” is for showing that the data displayed is for the
right probe.
3.
There are one
“Pass Word”
button and two sets of
“SET”/ “UP”/”DOWN”
buttons on the panel.
4.
There are no “on” and “off” buttons on the unit, the unit
works when the batteries are installed.
Alarm Sound Pattern (Same pattern for temperature high
/ low and probe unplugged alarms)
1 second (beep, beep) -> silence for 5 seconds -> 1
second (beep, beep) -> silence for 5 seconds …..
Temperatures are updated during the silent 5 seconds.
Alarm keeps sounding until it is turned off by pressing the
“PASSWORD”
button once. The function of the
“PASSWORD”
button is just for stopping the alarm when
the alarm sounds.
The key tone will sound when a key is pressed for a valid
function.
•
Battery cover removable with screwdriver only.
•
Unit to be water splash resistant (IPX4).
•
Velcro kit for mounting to outside unit.
1)
Temperature measuring range -40/105°C, -40/221°F
2)
Operating temperature range –20°C to 60°C / -4°F to 140°F
3)
Ac/- 2°F for -4°F to 212°F, +/-4°F for other ranges
4)
Power: 2 x AAA batteries
5)
Resolution 0.1°C / ºF
6)
Display: 2 lines readout
7)
Number of Probes = 2, one for each channel, each 6 feet in
length. Each probe works separately
8)
Probe design – PVC wire and stainless steel probe with PVC
material molded around the area joining the PVC wire and
the stainless steel probe. Temperature tolerance of the
probe wire is 105°C / 220°F
9)
Probe waterproof standard: (IPX7)
10)
Indicator – low battery
11)
“L”: for indicating the data is for Left probe
12)
“R”: for indicating the data is for Right probe
13)
Memory recall, provide ability to store at least 3 days and
maximum 4 days of alarming readings.
14)
All alarm event records will be kept in the memory even if
the battery dies. The user needs to change new batteries
when the icon of battery low appears.
15)
Total 120 alarm records, 60 alarm records per probe.
Recording Thermometer
with Dual Probes
-40º to 221ºF/-40º to 105ºC
R
ig
h
t
S
e
n
s
o
r
t
L
e
ft
S
e
n
s
o
r
s
SET
SET
UP
UP
DOWN
Pass
Word
DOWN
-40º to 221ºF/-40º to 105ºC
R
ig
h
t
S
e
n
s
o
r
t
L
e
ft
S
e
n
s
o
r
s
SET
SET
UP
UP
DOWN
Pass
Word
DOWN
General Operations
Alarm
Key Tone
Physical
Specifications
Password
System Reset
R 2 L
2
8
14
1
10
11
13
9
4
5
6
7
12
Password
System Reset
R 2 L
'AAA'
'AAA'
Figure# 1
Compartment Cover
Loosen but do not remove screws
Taylor® warrants this product to be free from defects in
material or workmanship for one (1) year from date of
original purchase. It does not cover damages or wear
resulting from accident, misuse, abuse, commercial use, or
unauthorized adjustment and/or repair. If service is required,
do not return to retailer. Should this product require service
(or replacement at our option), please pack the item carefully
and return it prepaid, along with store receipt showing the
date of purchase and a note explaining reason for return to:
Taylor Precision Products
2220 Entrada Del Sol
Las Cruces, New Mexico 88001
Customer Service Phone: 1-866-843-3905
There are no express warranties except as listed above. This
warranty gives you specific legal rights, and you may have
other rights which vary from state to state.
Made to our exact specifications in China.
© 2008 Taylor Precision Products and its affiliated companies,
all rights reserved. Taylor® and Leading the Way in
Accuracy® are registered trademarks of Taylor Precision
Products and its affiliated companies. All rights reserved.
One Year Limited Warranty
3